Agilent Liberty 100 Determination of Metals in Lubricating Oil by ICP-OES Application Note

This document describes the determination of 21 elements in an oil sample by ICP-OES.


Agilent Liberty 100 Determination of Metals in Lubricating Oil by ICP-OES Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
240 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
F341763300B1B0E3ACD3BBBBB80A66AC

About this manual

The document details a method for analyzing trace metals in lubricating oil to diagnose engine wear and contamination. It utilizes Inductively Coupled Plasma Atomic Emission Spectroscopy (ICP-OES) on an Agilent Liberty 100 instrument, featuring specific operating parameters optimized for organic solvents like kerosene.

The procedure involves a simple 1/10 weight/volume dilution of the oil sample with kerosene. The note provides detailed instrumental conditions, including gas flow rates and nebulizer pressure settings required to minimize carbon buildup when analyzing organics. It also lists detection limits for 21 specific elements and reports spike recovery percentages ranging from 92% to 104%, demonstrating the method's accuracy.

What's inside

  • Introduction
  • Instrumental Conditions
  • Sample Preparation
  • Standard Preparation
  • Analytical Conditions
  • Detection Limits

Specifications

RF Generator Frequency40.68 MHz
Monochromator Type0.75 m Czerny-Turner
Grating Specification1800 grooves/mm Holographic grating
Optical Resolution (1st order)0.018 nm
Plasma Gas Flow Rate12.0 L/min
Auxiliary Gas Flow Rate2.25 L/min
Torch TypeStandard one piece quartz torch with 1.4mm id injector tube
Nebulizer Pressure120 kPa
Pump Rate15 rpm
Sample Uptake Rate1.1 mL/min
Integration Time3 sec
PMT Voltage650 V

Frequently asked questions

What solvent is recommended for diluting oil samples?

Kerosene is commonly used, along with Xylene, DIBK, MIBK, or White spirit.

What is the maximum particle size suitable for this dilution method?

The dilution method is only applicable for wear metal particles up to 5 um.

What was the range of spike recoveries reported in this study?

The percentage spike recoveries ranged from 92% to 104%.