Agilent Vista-PRO ICP-OES Application Note Improving Throughput for Oils Analysis

This article describes how to improve sample throughput for ICP-OES oil analysis by improving the sample introduction system. By using a novel pump tubing arrangement, analysis time can be significantly reduced.


Agilent Vista-PRO ICP-OES Application Note Improving Throughput for Oils Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
52 KB
Published
12 June, 2012
Updated
28 September, 2023
MD5 checksum
394A88D9AB662825639AAB7411DF99D8

About this manual

The document details an optimization method for Inductively Coupled Plasma-Optical Emission Spectrometers (ICP-OES) specifically for analyzing wear metals in lubricating oils. The primary goal is to increase sample throughput by reducing the time required for sample introduction and washout, which are often the bottlenecks in analysis speed.

The proposed solution involves a novel pump tubing arrangement using an additional peristaltic pump tube connected via a T-piece. This setup directs excess sample flow to waste while maintaining high flow rates to the nebulizer, reducing sample uptake time by approximately 10 seconds and washout times significantly. The note provides specific instrument operating conditions, detection limits for various elements in kerosene, maximum measurable concentrations, and data demonstrating long-term stability over an eight-hour period.

What's inside

  • Introduction
  • Instrumental Setup
  • Detection Limits
  • Linear Range
  • Modified Pump Tubing Setup
  • Washout Performance
  • Long-Term Stability

Specifications

Plasma gas flow15.0 L/min
Auxiliary gas flow2.25 L/min
Nebulizer pressure or flow110 kPa or 0.60 L/min
Viewing height10mm
Pump speed12 rpm
Sample uptake delay15s
Stabilization time5s
Rinse time10s

Frequently asked questions

How much does the modified pump tubing setup reduce sample uptake time?

Using the modified setup, the sample uptake time was decreased by approximately 10 seconds.

What is the analysis speed achieved with this optimized system?

It is possible to accurately determine key wear metals and other elements in less than 50 seconds per sample.