Agilent Liberty 150 AX Turbo ICP-OES Application Note for Oil Analysis

This document describes how to use ICP-OES to determine 17 elements (Mo, Ag, Ti, Sn, Ba, B, Mn, Fe, Cr, Mg, V, Ca, Cu, Pb, Ni, Cd and Al) in a xylene solution of oil


Agilent Liberty 150 AX Turbo ICP-OES Application Note for Oil Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
239 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
30A5816C3241AF9BE20B9330BE2ED6EC

About this manual

The document describes a method to determine 17 specific trace elements (Mo, Ag, Ti, Sn, Ba, B, Mn, Fe, Cr, Mg, V, Ca, Cu, Pb, Ni, Cd, and Al) in a xylene solution of oil. The study addresses challenges such as plasma instability caused by volatile organic compounds and carbon deposition on the torch components.

What's inside

  • Introduction
  • Operation of the Ultrasonic Nebulizer/Membrane Desolvator
  • Experimental Equipment
  • Instrument Parameters
  • Calibration Solutions
  • Results
  • Conclusion

Specifications

RF Generator Power1.7 kW
Operating Frequency40.68 MHz
Monochromator Grating1800 grooves/mm holographic grating
Resolution (4th order)0.006 nm

Frequently asked questions

Which elements can be determined in oil using this method?

The method determines 17 elements: Mo, Ag, Ti, Sn, Ba, B, Mn, Fe, Cr, Mg, V, Ca, Cu, Pb, Ni, Cd, and Al.

What is the recovery range for trace elements in oil?

Recoveries for the analyzed sample varied from 89% to 107%.