Agilent Vista Axial ICP-OES Environmental Analysis Application Note

Agilent Vista ICP-OES is a unique instrument that can simultaneously measure the entire elemental spectrum. It has distinct advantages that can be used to analyze difficult samples with a maximum of 5% total dissolved solids and eliminate cooler tails of plasma, reduce easily ionizable element (EIE) interferences and maximize linear dynamic range.


Agilent Vista Axial ICP-OES Environmental Analysis Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
14
File format
PDF
File size
348 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
A46B674C24A43389E810A9932C889840

About this manual

The document presents a complete analytical method developed on the Agilent Vista simultaneous ICP-OES system with axially viewed plasma. Key features highlighted include the VistaChip CCD, RF robustness allowing analysis of difficult samples up to 5% total dissolved solids, and the Cooled Cone Interface (CCI) which reduces interferences.

The note details the optimization of instrument operating conditions and software settings such as MultiCal and Adaptive Integration to meet US EPA data quality control protocols. It provides specific instrument parameters, detection limits for various elements, and discusses the advantages of axial viewing over dual-viewed systems in terms of analysis time and cost.

What's inside

  • Introduction
  • Instrument Set up
  • Ionization Buffer
  • Method Detection Limits (MDL)
  • Linear Dynamic Range (LDR)
  • US EPA Guidelines
  • VistaChip Features
  • Cooled Cone Interface

Specifications

Power1.40 kW
Plasma gas flow15.0 L/min
Auxiliary gas flow0.75 L/min
Pump speed15 rpm

Frequently asked questions

What are the advantages of using axial viewing on the Agilent Vista?

The RF robustness permits analysis of difficult samples up to 5% total dissolved solids, and it eliminates the cooler tail of the plasma to reduce interferences.

How does MultiCal software assist in dynamic range coverage?

It allows multiple wavelengths to be used simultaneously for the same element, automatically assigning sample results to appropriate wavelengths within the valid linear dynamic range.

Which US EPA guidelines are referenced in this method?

The primary guiding documents used are CLP ILM0 4.0, ILM05.0, Methods 200.7 and 6010B.