Agilent 7700x ICP-MS Application Note for FGD Wastewater Analysis

This document is an application note on the analysis of flue gas desulfurization wastewaters using Agilent 7700x ICP-MS. It discusses the U.S. Environmental Protection Agency's (USEPA) revision of effluent guidelines for the steam electric power generating industry, specifically addressing the increased wastewater discharges from Phase 2 of the Clean Air Act amendments. The application note highlights the measurement challenges of toxic metals in FGD wastewater and introduces the application of Agilent 7700x ICP-MS.


Agilent 7700x ICP-MS Application Note for FGD Wastewater Analysis - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
188 KB
Published
11 June, 2012
Updated
29 September, 2023
Model
7700x
MD5 checksum
7B3155A6F4A403162D994A56691ABFC9

About this manual

The document describes a method developed to analyze regulated metals in high-matrix flue gas desulfurization (FGD) wastewaters. The U.S. Environmental Protection Agency (USEPA) requires these measurements due to revised effluent guidelines for steam electric power plants. FGD wastewater presents challenges due to high dissolved solids and potential matrix-based polyatomic ion interferences.

The Agilent 7700x ICP-MS system, equipped with the High Matrix Introduction (HMI) system, Octopole Reaction System (ORS), and optional ISIS-DS discrete sampling system, was used to develop a robust analytical method. The method includes specific sample preparation protocols, instrument parameters, and quality control procedures to ensure accurate detection of toxic metals in this challenging matrix.

What's inside

  • Introduction
  • Instrumentation
  • Sample preparation
  • Analytical method
  • Quality control
  • Results

Specifications

Forward RF power1550 W
Carrier gas flow0.56 L/min
Dilution gas flow0.33 L/min
Sample loop volume600 uL
Online dilution factor1:2

Frequently asked questions

Which systems are used for the FGD wastewater analysis method?

The method uses an Agilent 7700x ICP-MS equipped with an Agilent ISIS-DS discrete sampling system.

How does the HMI system help with sample analysis?

The High Matrix Introduction (HMI) system enables controlled, reproducible aerosol dilution, which increases plasma robustness and reduces exposure to the sample matrix.

What is the purpose of the ORS in this method?

The Octopole Reaction System (ORS) operating in helium collision mode eliminates matrix-based polyatomic interferences regardless of sample composition.

Download PDF (188 KB)

More Agilent Manuals