Agilent 5990-7765EN Application Note: Use Temperature to Enhance Oligonucleotide Mass Transfer

This application note describes the use of temperature to enhance oligonucleotide mass transfer and improve resolution in ion-pair reversed phase HPLC. Temperature enhancement of mass transfer can improve oligonucleotide separation resolution, enabling the use of smaller particle columns and reduced column lengths. Temperature enhancement of resolution is particularly useful for oligonucleotide analysis, as oligonucleotides have relatively low retention in reversed phase chromatography. Temperature also accelerates oligonucleotide separations, increasing laboratory throughput.


Agilent 5990-7765EN Application Note: Use Temperature to Enhance Oligonucleotide Mass Transfer - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
741 KB
Published
11 June, 2012
Updated
01 October, 2023
MD5 checksum
A8D120F7DF6CC15BB2EC2D2EE2AD0FBE

About this manual

This document explores the use of elevated temperatures to enhance oligonucleotide mass transfer and improve resolution in ion-pair reversed phase HPLC. It addresses the limitations of traditional silica-based columns at high temperatures and compares their performance against hybrid and polymeric columns.

What's inside

  • Introduction
  • Materials and Methods
  • Temperature dependence of oligonucleotide separations
  • Results and Discussion
  • Conclusions
  • References

Specifications

Column A TypePLRP-S 100A, 3 ym, PS/DVB HPLC column
Column B Type2.5 um, 125A Hybrid silica-methylsiloxane C18 column
Column C Type1000A, 3 ym, Thermally stable bonded silica C18 column
Column D Type100A, 5 um, Traditional silica C18 column
Eluent A Composition100 mM Triethylammonium acetate (TEAA) in HPLC-grade water
Eluent B Composition100 mM TEAA in 25:75 Acetonitrile:water
Flow Rate1.0 mL/min or 0.45 mL/min
Detection WavelengthUV, 254 nm

Frequently asked questions

Which column performs best at 80 °C for oligonucleotide separation?

The PLRP-S column provides superior performance at 80 °C compared to other columns.

What is the primary benefit of using elevated temperatures in this analysis?

Elevated temperatures improve the mass transfer of denatured oligonucleotides between the stationary and mobile phases, enhancing resolution.