Agilent 6410 LC/MS Application Note: Determination of Estrone-3-Sulfate in Water

This application describes the development of a liquid chromatography/tandem mass spectrometry (LC-MS/MS) method for the determination of estrone-3-sulfate (E1S) in water with a sensitivity below the ppt level (< ng L-1). The method involves solid-phase extraction (SPE) and liquid chromatography/tandem mass spectrometry (LC-MS/MS) in negative electrospray mode. The method achieved high recoveries (>93%) and low detection limits (<1 pg L-1). In addition, the method was highly selective, with no interferences from other compounds observed. This makes it a suitable method for the analysis of E1S in environmental water samples.


Agilent 6410 LC/MS Application Note: Determination of Estrone-3-Sulfate in Water - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
180 KB
Published
08 June, 2012
Updated
30 September, 2023
MD5 checksum
49515B975AB45479E2CE0B5A2E3598EB

About this manual

This document details an analytical method for quantifying estrone-3-sulfate (E1S) in water samples at ultra-trace levels (sub-ppt). The method utilizes solid phase extraction (SPE) for sample preparation followed by liquid chromatography-tandem mass spectrometry (LC-MS/MS) in negative electrospray mode.

The analysis is performed using an Agilent 1200 Infinity LC system coupled with an Agilent 6410 Triple Quadrupole Mass Spectrometer. The method employs a Gemini C18 column and monitors specific MRM transitions for both the analyte and an internal standard (boldenone-sulfate-d3) to ensure high specificity and sensitivity.

What's inside

  • Sample Preparation Procedure
  • Calibration Curve
  • LC-MS/MS Measurement
  • MS/MS Parameters
  • Analytical Performance
  • Validation Parameters
  • Analysis of Unknown Samples
  • Conclusions

Specifications

ColumnGemini (50 mm x 2 mm, 3 µm) C18 110A
Column temperature40°C
Flow rate0.3 mL/min
Capillary voltage3500 V
Nebulizer pressure40 psi
Drying gas flow13 L/min
Gas temperature275°C
Injected volume10 µL

Frequently asked questions

What is the detection limit for estrone-3-sulfate in this method?

The method analyzes estrone-3-sulfate down to sub-ppt level, specifically below 1 ng/L.

Which internal standard is used for this analysis?

The method uses boldenone-sulfate-d3 as the internal standard.

What are the MRM transitions monitored for estrone-3-sulfate?

The method monitors transitions at 349.2 > 269.2 and 349.2 > 146.2.