Agilent 5975T LTM GC/MS Application Note for Drug Detection


Agilent 5975T LTM GC/MS Application Note for Drug Detection - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
636 KB
Published
08 June, 2012
Updated
08 June, 2012
MD5 checksum
5055C5D2A750F89B0EBA42AE47E4A11A

About this manual

This document details a rapid, low-cost method for the qualitative detection of 63 drugs of abuse in whole blood samples. The method utilizes the innovative Thermal Separation Probe (TSP) sample introduction technique combined with the Low Thermal Mass (LTM) technology on the Agilent 5975T GC/MS system. A key advantage is that no sample cleanup is required, as non-volatile components remain in the TSP micro-vial after volatilization. The method relies on Agilent Deconvolution Reporting Software (DRS) to streamline data handling and identify compounds within the complex blood matrix.

The note presents experimental results demonstrating the method's sensitivity and speed. Using a fast temperature gradient, the analytical run time is reduced to approximately 6.5 minutes compared to over 28 minutes for conventional methods. The study confirms that over 94% of the 56 thermally stable drugs analyzed achieved a method detection limit of 5 ng/pL. The document includes chromatograms and deconvolution examples, such as the identification of chlorpromazine in a blood sample from a fatal overdose subject.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Results and Discussion
  • Drug Detection in Whole Blood
  • Method Confirmation
  • Conclusion

Specifications

Analytical columnAgilent HP-5ms LTM 10 m x 0.18 mm, 0.18 µm
Injection volume3-5 µL
Inlet temperature280 °C
Ion source temperature230 °C
Method detection limit5 ng/pL

Frequently asked questions

Does this method require sample cleanup?

No, no sample cleanup is required as non-volatile components remain inside the TSP micro-vial after the drug compounds are volatilized.

What software is used for compound identification?

Agilent Deconvolution Reporting Software (DRS) is used for compound identification.