Agilent 1260 Infinity Hybrid SFC/UHPLC System Application Note: Transfer of USP Prednisolone Method from HPLC to SFC

This Application Note describes the transfer of a United States Pharmacopeia (USP) method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System. The results show that the SFC method meets the system suitability criteria, is 4x faster, and results in 17x lower solvent expenses. Robustness tests on the SFC method demonstrate excellent robustness for routine analysis.


Agilent 1260 Infinity Hybrid SFC/UHPLC System Application Note: Transfer of USP Prednisolone Method from HPLC to SFC - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
765 KB
Published
08 June, 2012
Updated
29 September, 2023
MD5 checksum
6C2CD38EAE3A3CEC7DA5359219A94290

About this manual

The document details a method transfer for analyzing prednisolone, replacing a toxic and expensive normal phase HPLC method with a faster, eco-friendly SFC method. The study utilizes the Agilent 1260 Infinity Hybrid SFC/UHPLC System to perform both methods on a single instrument, eliminating system variability and saving costs.

What's inside

  • Introduction
  • Experimental Instruments
  • Chromatographic parameters
  • Preparation of standards
  • Linearity and robustness sample preparation
  • Procedure
  • Results and discussion

Frequently asked questions

What is the SFC flow rate used for the prednisolone method?

The SFC flow rate is maintained at 2.9 mL/min.

What is the run time for the SFC method?

The SFC run time is 5.5 minutes.