This document provides information about Agilent Technologies' 7696A WorkBench Sample Preparation Station, including the features of the equipment and usage warnings.
This document outlines the necessary site preparation steps for installing the Agilent 7696A Workbench system. It covers the client's responsibilities, required space dimensions, and necessary supplies such as flasks and solvents. This manual is written in French.
| Height | 70.4 cm (27.72 in) |
|---|---|
| Power Consumption (Europe) | 800 VA |
A total of 174 cm (68 in) of free space on the bench is required for a complete CPG system, considering access and the printer.
You must prepare the workbench space before the system arrives, specifically ensuring a total height of at least 70.4 cm (27.72 in) is available.
A complete CPG system requires a total of 174 cm (68 in) of clear space on the workbench.