Agilent MMI Large Volume Injection Tutorial

Large volume injection (LVI) methods can improve the sensitivity and selectivity of existing analytical methods. This tutorial introduces the LVI capabilities of the Agilent Multimode Injector (MMI), including Hot Splitless Injection, Cold Splitless Injection, and Solvent Vent.


Agilent MMI Large Volume Injection Tutorial - cover page
Brand
Agilent
Category
Other Electronics
Document type
User Manual
Language
French
Pages
23
File format
PDF
File size
2.7 MB
Published
08 June, 2012
Updated
30 September, 2023
MD5 checksum
BF7B6710ED28F7823F364C382231EDE0

About this manual

This document is a tutorial for the Agilent Multimode Injector (MMI), focusing on Large Volume Injection (LVI) techniques to improve trace analysis. It explains the benefits of LVI, such as increased sensitivity and reduced sample consumption, and details the challenges associated with traditional hot splitless injection, including analyte degradation and needle fractionation. The text introduces the MMI's programmable temperature capabilities as a solution to these issues. This manual is written in French.

What's inside

  • Injection chaude sans division
  • Injection froide sans division
  • Mode Solvent Vent
  • Principe de l'injection à grand volume
  • Avantages de la LVI
  • Défis de l'injection chaude
  • Configuration de la méthode
  • Paramètres de l'injecteur MMI

Specifications

Sample Volume Rangea few microliters to at least one milliliter
Injector Temperature (Hot Splitless)250 °C
Column Flow Rate4ml/min
Syringe Capacity10 pl

Frequently asked questions

What are the benefits of Large Volume Injection (LVI)?

LVI improves trace analysis by increasing the mass of the analyte reaching the detector, leading to higher signal-to-noise ratios and lower detection limits.

What is the maximum injection volume supported by Large Volume Injection methods?

Large Volume Injection can be applied to volumes ranging from a few microlitres up to at least one millilitre.

What is the nominal volume of standard injector liners mentioned in the text?

Most standard injector liners have a nominal volume of one millilitre.