Introduces the use of Agilent G1674AA Deconvolution Reporting Software (DRS) Solution for Forensic Toxicology.
This manual guides users through setting up and using the Agilent G1674AA Deconvolution Reporting Software (DRS) for forensic toxicology. It covers method naming conventions, database library setup, system verification, chromatographic considerations, and calibration steps. The guide includes detailed tables for different column types and analysis speeds, helping users configure their systems based on hardware and application needs.
The manual is structured to help users choose appropriate method files based on stationary phase, analysis speed, and instrument configuration. It provides step-by-step instructions for setting up the DRS system, including retention time calibration and response factor calibration, and details how to integrate with MS ChemStation for post-run analysis.
Method file names begin with FT, for Forensic Toxicology.
DB-5MS method sets are typically the better choice because runs end at a lower temperature.
Method sets 3X, 4X, and 6X require the fast oven (240 V).