Agilent 6460 Triple Quad LC/MS System Quick Start Guide
This document provides a quick start guide for the Agilent 6460 Triple Quad LC/MS system, including steps for starting the data acquisition software, preparing the LC modules and Triple Quad instrument, setting up and running acquisition methods, reviewing results with the Qualitative Analysis program, and setting up and running quantitation. The Agilent 6460 Triple Quad LC/MS is a liquid chromatograph-triple quadrupole mass spectrometer that performs MS/MS using three sets of parallel rods (quadrupole, hexapole, quadrupole). The first quadrupole separates ions into precursor ions that are fragmented in the hexapole into product ions, which are separated by the second quadrupole. Often, two or more precursor ions and their product ions are monitored in sequence in MRM (multiple reaction monitoring) mode. The Agilent 6460 Triple Quad LC/MS is shipped with the Agilent Jet Stream Technology, which utilizes a super-heated sheath gas to collimate the nebulizer spray and increase the number of ions entering the mass spectrometer. The system can be set up with either the Agilent 1100LC or 1200LC modules. It also includes the Agilent MassHunter Workstation Software with three programs.
This guide introduces the Agilent 6460 Triple Quad LC/MS System, a liquid chromatograph‑triple quadrupole mass spectrometer that uses Jet Stream Technology. It provides a roadmap for new users to get the instrument up and running, covering software, hardware, and workflow basics.
The guide is organized into six main steps: start the Data Acquisition software, prepare the LC modules, prepare the Triple Quad instrument, set up and run an acquisition method, review results with Qualitative Analysis, and set up and run quantitation. Each step includes detailed screen instructions, pane descriptions, and tips for successful operation.
What's inside
Getting Started
Start the Data Acquisition software
Prepare the LC modules
Prepare the Triple Quad instrument
Set up and run an acquisition method
Review results with Qualitative Analysis
Set up and run quantitation
Specifications
Mass range
2 to 3000 amu
Maximum MRMs per time segment
up to 500
Polarity switching delay
100 ms
LC flow rate
800 µL/min
Nebulizer pressure
35 psi
Frequently asked questions
How do I start the Data Acquisition software?
Double‑click the Data Acquisition icon on the desktop to launch the program.