This manual provides detailed information on the features and usage of the Agilent 1200 Series Standard and Preparative Autosamplers.
This manual provides instructions for using Agilent 1200 Series autosamplers, including standard and preparative models. It covers setup, sample handling, performance optimization, maintenance, and troubleshooting. The document includes detailed sections on site requirements, installation procedures, sample trays, injection volumes, and maintenance functions.
The autosamplers are designed for use with Agilent 1200 Series LC systems or compatible systems with remote control. Three models are described: G1329A (standard), G1329B (standard SL), and G2260A (preparative). The document also details cable connections, safety notices, and maintenance kits for each model.
| Injection Volume (Analytical Head) | 0.1 - 100 µL |
|---|---|
| Injection Volume (Preparative Head) | 0.1 - 900 µL |
| System Pressure Limit (Preparative Head) | 200 bars or 400 bars |
| Maximum Pressure (G1329B) | 600 bar |
| Sample Rack Sizes | Full-size (100 x 1.8 ml vials), Half-size (40 x 1.8 ml or 15 x 6 ml vials) |
Three models are available: G1329A (standard), G1329B (standard SL), and G2260A (preparative).
The preparative head provides injection volumes from 0.1 to 900 µL.
The standard analytical head provides injection volumes from 0.1 to 100 µL.
Full-size rack (100 x 1.8 ml vials), half-size rack (40 x 1.8 ml vials), and half-size rack (15 x 6 ml vials).