This document describes the calibration standard procedures for the GOODWILL INSTRUMENT GFG-8216A Function Generator
This document provides the internal calibration procedures for the Goodwill Instrument GFG-8216A function generator. It outlines each calibration station, the required adjustments, test equipment, and acceptance criteria to ensure the instrument meets its performance specifications.
The manual is organized into six calibration stations covering initial VR settings, puncture testing, functional checks, voltage and frequency adjustments, distortion measurements, and final verification of output waveforms. Each step includes specific measurement values, equipment settings, and remarks for the technician.
Adjust all VRs to MAX.