Agilent 5500 series FTIR user manual, introduces the features and characteristics of the product, including connecting power, connecting computer, opening instrument, starting software, etc.
This manual describes how to unpack, install, and operate the Agilent 5500 series FTIR spectrometers. It covers power and computer connections, software startup, performance verification, and detailed procedures for analyzing liquids, powders, gases, and pastes using the TumblIR, ATR, and gas‑cell accessories.
Safety warnings, recommended accessories, spare‑part numbers, and full technical specifications are included. The document also lists optional accessories and provides guidance for maintaining the instrument in laboratory environments. This manual is written in Chinese.
| Interferometer geometry | High‑throughput Michelson interferometer with fixed and moving plane mirrors |
|---|---|
| Standard beamsplitter | Zinc selenide |
| Maximum spectral resolution | 4 cm⁻¹ |
| Laser | Low‑power solid‑state laser |
| Light source | Wire‑wound element |
| Spectral range | 4000 to 650 cm⁻¹ |
| Detector | 1.3 mm diameter, hot‑cooled dTGS |
| Power supply | 100/120/240 V AC, 3 A, 50‑60 Hz |
| Operating temperature range | -10 °C to 50 °C |
It requires 100/120/240 V AC, 3 A, at 50‑60 Hz.