Canberra 9233653E V3.0 Error #1144: DETECTOR IS BEHIND THE ABSORBER.

Genie™ 2000 Spectroscopy Software · taken from the manufacturer’s manual, pages cited

Error code#1144

The detector location conflicts with the generic absorber thickness defined in the 'beaker' template. The manual's first action is: “Change the parameters of the source/detector location or/and the total thickness of the used generic absorbers.”

Model
Canberra 9233653E V3.0
Display shows
DETECTOR IS BEHIND THE ABSORBER. (#1144)
Type
Error code
Source
Manual p. 432

What #1144 means

DETECTOR IS BEHIND THE ABSORBER. p. 432

Common causes

  • A conflict was detected between the detector location defined in the 'beaker' template and the generic absorber thickness. p. 432

What to do

  1. Change the parameters of the source/detector location or/and the total thickness of the used generic absorbers. p. 432

Before you service the product, read the safety instructions in the manual p. 526.

If #1144 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Canberra service.

This page summarises what the manufacturer’s manual says about #1144; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Canberra. Found a mistake? Report an error.