Two fragments in the beaker description coincide completely, which is not allowed for boundary or inner/outer contours. The manual's first action is: “Neither an inner nor an outer fragment is allowed to fully coincide with a boundary fragment. Change the erroneous fragment coordinates to fix the error. Then try again.”
- Model
- Canberra 9233653E V3.0
- Display shows
- #1124
- Type
- Error code
- Source
- Manual p. 430
What #1124 means
THE BEAKER DESCRIPTION FILE CONTAINS TWO COINCIDING FRAGMENTS. A BOUNDARY FRAGMENT IS ONE OF THEM. p. 430
Common causes
- THE BEAKER DESCRIPTION FILE CONTAINS TWO COINCIDING FRAGMENTS. A BOUNDARY FRAGMENT IS ONE OF THEM. p. 430
What to do
- Neither an inner nor an outer fragment is allowed to fully coincide with a boundary fragment. Change the erroneous fragment coordinates to fix the error. Then try again. p. 430
Before you service the product, read the safety instructions in the manual p. 526.
If #1124 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Canberra service.
This page summarises what the manufacturer’s manual says about #1124; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Canberra. Found a mistake? Report an error.