instruction manual
This manual provides detailed instructions for installing and handling the Endevco 7264B piezoresistive accelerometer. It includes material requirements, torque specifications, and precautions to avoid damage. The document also covers initial check-out procedures such as impedance and insulation resistance tests.
Calibration and maintenance are addressed, including recalibration intervals and cleaning instructions. Users are advised to contact Endevco for support or recalibration if needed.
Use 2.6 in-lbf (0.29 N-m) recommended, 3.0 in-lbf (0.34 N-m) maximum.
Wipe clean with a damp cloth and solvent such as acetone. Do not soak or immerse the unit in any solvent or water.
Sensitivity and ZMO calibration should be performed every 6 to 12 months, depending on usage.