Key facts
- Brand
- Applied Biosystems
- Model
- ABI Prism 7900HT
- Category
- medical technology
- Documents
- 3
- Languages
- ZH · EN
- Last added
- 2023-11-10
Start here

ABI PRISM 7900HT Sequence Detection System User Guide
This document serves as the official User Guide for the ABI PRISM 7900HT Sequence Detection System. It provides detailed instructions on getting started with the hardware and software, including system setup and basic operational procedures. The guide covers the use of the Sequence Detection System (SDS) software for managing data and running various PCR applications.
Read onlineApplied Biosystems ABI Prism 7900HT specifications
| Heated cover temperature | 105 °C |
|---|---|
| Standard reaction volume, 96-well plate | 50 μL |
| Standard reaction volume, 384-well plate | 10 μL |
| Standard reaction volume, micro card | 2 μL |
| Recommended minimum volume, 96-well plate | greater than 25 μL |
| Recommended minimum volume, 384-well plate | greater than 5 μL |
| Default thermal profile | 50 °C 2 min; 95 °C 10 min; 40 cycles of 95 °C 15 sec and 60 °C 1 min |
| Computer login password (Administrator) | 7900 |
Source: Applied Biosystems ABI Prism 7900HT Sequence Detection System Quick Reference Guide
All Applied Biosystems ABI Prism 7900HT documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
热盖需要加热到多少度?
电源打开后主机开始加热热盖,直到 105 °C;若未达到该温度就开始实验,仪器会自动暂停等待。
From: Applied Biosystems ABI Prism 7900HT Sequence Detection System Quick Reference Guide
融解曲线试验适用于哪种方法?
融解曲线试验只适用于 SYBR Green I 荧光染料法,不适用于荧光探针法。
From: Applied Biosystems ABI Prism 7900HT Sequence Detection System Quick Reference Guide
What is the product name and its stated use?
The ABI PRISM 7900HT is a Sequence Detection System for PCR processes.
What is the stated licensing requirement for PCR use?
A license is required to practice the PCR process.
What is the standard reaction volume for a micro reaction card?
The standard reaction volume for a micro reaction card is 2 μL.
From: Applied Biosystems ABI Prism 7900HT Fluorescent Quantitative PCR System User Manual
To which temperature must the thermal lid heat up before the experiment starts?
The thermal lid must heat up to 105°C before the experiment begins.
From: Applied Biosystems ABI Prism 7900HT Fluorescent Quantitative PCR System User Manual