Applied Biosystems 3130/3130 xl Manuals

ManualLib has 4 documents for the Applied Biosystems 3130/3130 xl (Analyzers): 2 user manuals, 1 product leaflet, 1 quick start guide. Every file can be read online or downloaded as a PDF for free.

User Manual 2Product Leaflet 1Quick Start Guide 1

Key facts

Model
3130/3130 xl
Category
Analyzers
Industry
Health
Documents
4
Languages
EN
Last added
2026-11-18

Start here

Applied Biosystems 3130/3130xl Genetic Analyzers Quick Reference Card

Applied Biosystems 3130/3130xl Genetic Analyzers Quick Reference Card

Quick Start Guide 3 pages · 420 KB · EN

This quick reference card provides essential operational and maintenance instructions for the Applied Biosystems 3130/3130xl Genetic Analyzers. It covers daily and weekly maintenance tasks, instrument setup, calibration procedures, and software navigation. The guide supplements the Getting Started and Maintenance manuals and includes safety information and run module specifications for different sequencing applications.

Key tasks include checking reservoirs and polymer levels, cleaning surfaces and fittings, and performing spatial and spectral calibrations. It also details run module specifications such as capillary length, polymer type, run time, and throughput for various analysis types including fragment analysis and SNP detection.

Read online

Applied Biosystems 3130/3130 xl specifications

Capillary Number (3130)4-capillary
Capillary Number (3130 xl)16-capillary
Polymer DeliveryAutomated pump system
Software WizardsYes (Install Array, Change Polymer Type, Replenish Polymer, Bubble Remove, Water Wash, Shutdown, Auto-sampler Calibration, Update Cap Array Info)
Data Collection Softwarev3.0
Operating SystemWindows 2000/XP

Source: Applied Biosystems 3130/3130 xl Genetic Analyzer Product Leaflet

All Applied Biosystems 3130/3130 xl documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Quick Start Guide 1

The shortest path to a working setup, usually only a few pages.

Questions answered in these manuals

How often should buffer and water levels be checked?

Ensure adequate levels of buffer and water in the reservoirs before each run.

From: Applied Biosystems 3130/3130 xl Genetic Analyzer Maintenance, Troubleshooting, and Reference Guide

How often should polymer be replaced?

Replace the polymer using the Replenish Polymer Wizard weekly or as needed.

From: Applied Biosystems 3130/3130 xl Genetic Analyzer Maintenance, Troubleshooting, and Reference Guide

Why should I upgrade from a 310 to a 3130 or 3130 xl Genetic Analyzer?

Upgrading provides increased throughput allowing more samples to be processed with less hands-on time. The novel polymer delivery system and software wizards increase ease of use and minimize maintenance time. Both instruments are compatible with GeneMapper ID Software v3.2 and AmpFLSTR kits.

From: Applied Biosystems 3130/3130 xl Genetic Analyzer Product Leaflet

Is the 3130 xl Genetic Analyzer validated for use for forensic casework and database samples?

Yes, Applied Biosystems has determined that the 3130 series instruments produce robust, reliable, reproducible, and accurate results when used with AmpFLSTR kits for genomic DNA analysis per DAB/SWGDAM guidance. However, laboratories should perform their own validation studies.

From: Applied Biosystems 3130/3130 xl Genetic Analyzer Product Leaflet

What is the maintenance schedule for the 3130/3130xl instrument?

Daily tasks include checking reservoirs, removing bubbles, and cleaning surfaces. Weekly tasks include cleaning pump blocks and checking storage conditions. Monthly tasks include defragmenting the hard drive and replacing polymer older than one week.

From: Applied Biosystems 3130/3130xl Genetic Analyzers Quick Reference Card

What are the run module specifications for the 3130xl?

The 3130xl supports various run modules including Ultra rapid (36 cm, POP-4®, 40 min, 576 samples), Rapid (36 cm, POP-6™, 60 min, 384 samples), and Fast (50 cm, POP-7™, 60 min, 384 samples).

From: Applied Biosystems 3130/3130xl Genetic Analyzers Quick Reference Card