The ABI Prism® 7900HT Sequence Detection System is a high-throughput, high-sensitivity fluorescence quantitative PCR instrument for gene expression analysis, SNP genotyping, microbial detection and other applications.
This Chinese-language concise manual for the Applied Biosystems ABI Prism 7900HT real-time PCR system walks through the SDS software workflows used in the laboratory. It starts with a quick-start section covering powering up the computer and the 7900HT main unit, creating a new plate file, defining detectors, entering the sample table, setting the PCR run and analyzing results.
The remaining sections give step-by-step procedures for three assay types: absolute quantification with a standard curve, relative quantification, and end-point plate reading. They describe instrument warm-up (the heated cover rises to 105 °C), detector and marker setup, thermal profile editing including the dissociation stage, reaction volumes, passive reference selection, saving templates, and viewing amplification plots, standard curves and reports. This manual is written in Chinese.
| Heated cover temperature | 105 °C |
|---|---|
| Standard reaction volume, 96-well plate | 50 μL |
| Standard reaction volume, 384-well plate | 10 μL |
| Standard reaction volume, micro card | 2 μL |
| Recommended minimum volume, 96-well plate | greater than 25 μL |
| Recommended minimum volume, 384-well plate | greater than 5 μL |
| Default thermal profile | 50 °C 2 min; 95 °C 10 min; 40 cycles of 95 °C 15 sec and 60 °C 1 min |
| Computer login password (Administrator) | 7900 |
电源打开后主机开始加热热盖,直到 105 °C;若未达到该温度就开始实验,仪器会自动暂停等待。
融解曲线试验只适用于 SYBR Green I 荧光染料法,不适用于荧光探针法。
使用 ABI 的 TaqMan Universal PCR Master Mix 或 SYBR Green PCR Master Mix 时选 ROX;试剂中没有参比荧光则选 None。
标准反应体积为 50 μL,如需修改建议 96 孔板大于 25 μL,且同一块板上所有孔体积必须一致。