Applied Biosystems ABI Prism 7900HT Sequence Detection System Quick Reference Guide

The ABI Prism® 7900HT Sequence Detection System is a high-throughput, high-sensitivity fluorescence quantitative PCR instrument for gene expression analysis, SNP genotyping, microbial detection and other applications.


Applied Biosystems ABI Prism 7900HT Sequence Detection System Quick Reference Guide - cover page
Brand
Applied Biosystems
Category
medical technology
Document type
User Manual
Language
Chinese
Pages
17
File format
PDF
File size
1.2 MB
Published
10 November, 2023
Updated
10 November, 2023
MD5 checksum
C229ECD825D746D7246BC9451B37236F

About this manual

This Chinese-language concise manual for the Applied Biosystems ABI Prism 7900HT real-time PCR system walks through the SDS software workflows used in the laboratory. It starts with a quick-start section covering powering up the computer and the 7900HT main unit, creating a new plate file, defining detectors, entering the sample table, setting the PCR run and analyzing results.

The remaining sections give step-by-step procedures for three assay types: absolute quantification with a standard curve, relative quantification, and end-point plate reading. They describe instrument warm-up (the heated cover rises to 105 °C), detector and marker setup, thermal profile editing including the dissociation stage, reaction volumes, passive reference selection, saving templates, and viewing amplification plots, standard curves and reports. This manual is written in Chinese.

What's inside

  • 快速入门
  • 开机
  • 新建文件
  • 探针设置
  • 参数设置
  • 填样品表
  • 启动循环
  • 数据分析
  • 查看结果
  • 绝对定量

Specifications

Heated cover temperature105 °C
Standard reaction volume, 96-well plate50 μL
Standard reaction volume, 384-well plate10 μL
Standard reaction volume, micro card2 μL
Recommended minimum volume, 96-well plategreater than 25 μL
Recommended minimum volume, 384-well plategreater than 5 μL
Default thermal profile50 °C 2 min; 95 °C 10 min; 40 cycles of 95 °C 15 sec and 60 °C 1 min
Computer login password (Administrator)7900

Frequently asked questions

热盖需要加热到多少度?

电源打开后主机开始加热热盖,直到 105 °C;若未达到该温度就开始实验,仪器会自动暂停等待。

融解曲线试验适用于哪种方法?

融解曲线试验只适用于 SYBR Green I 荧光染料法,不适用于荧光探针法。

参比荧光(Passive Reference)应该怎么选?

使用 ABI 的 TaqMan Universal PCR Master Mix 或 SYBR Green PCR Master Mix 时选 ROX;试剂中没有参比荧光则选 None。

96 孔板反应体积可以设多大?

标准反应体积为 50 μL,如需修改建议 96 孔板大于 25 μL,且同一块板上所有孔体积必须一致。