Applied Biosystems PRISM 377 DNA Sequencer 96-Lane Upgrade User's Manual
This document serves as a user's manual for the ABI PRISM 377 DNA Sequencer 96-lane Upgrade Kit. It details the enhancements provided by the kit, which allow the instrument to support up to 96 lanes for both Sequencing and GeneScan applications. Key features include an increased scan window, improved data collection capabilities, and upgraded hardware components like combs and clamps.
Read onlineApplied Biosystems PRISM 377 specifications
| Max Lanes | 96 lanes |
|---|---|
| Data Collection Channels | 480 channels |
| Comb Thickness | 0.4 mm |
| Minimum Gel Thickness | 0.2 mm |
| Max Loading Volume | 1.5 µL |
| Clamping Pressure | 10–12 lbs |
| Compatible Instrument | 377-36 or 377XL DNA Sequencer |
| Required OS | Mac OS 8 |
Source: Applied Biosystems PRISM 377 DNA Sequencer 96-Lane Upgrade User's Manual
All Applied Biosystems PRISM 377 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What combs are supported for 96-lane runs?
The 96-lane run mode only supports the use of a shark's-tooth comb.
From: Applied Biosystems PRISM 377 DNA Sequencer 96-Lane Upgrade User's Manual
Does the processor speed affect gel collection?
No, the processor speed does not impact Collection, but it does impact the speed of analysis.
From: Applied Biosystems PRISM 377 DNA Sequencer 96-Lane Upgrade User's Manual
What is the primary enhancement provided by the XL upgrade?
The primary enhancement is an increase in the number of samples that can be analyzed simultaneously.
What run mode must be used for 48, 50, 64, and 66 lanes?
You must use the XL Scan Run mode for these lane configurations.