Applied Biosystems QuantStudio 6 Pro and 7 Pro Real-Time PCR Systems User Guide
This document is the User Guide for the Applied Biosystems QuantStudio 6 Pro and QuantStudio 7 Pro Real-Time PCR Systems. It covers network and password security requirements, instrument hardware descriptions, and an overview of data collection features. The guide details the use of VeriFlex Zones, AutoDelta settings, and hands-free features such as facial authentication and voice commands. It also describes the Thermo Fisher Connect platform and its integration with the instrument and desktop software.
Read onlineAll Applied Biosystems 7 Pro documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the maximum temperature difference allowed between adjacent areas when setting gradient reaction temperatures?
The temperature difference between adjacent areas must not exceed 5 degrees Celsius.
From: Applied Biosystems QuantStudio 6 & 7 Pro Real-Time PCR Instrument User Manual
What is the valid temperature range for the Pause Cycle setting?
The pause temperature can be set anywhere between 4 and 99.9 degrees Celsius.
From: Applied Biosystems QuantStudio 6 & 7 Pro Real-Time PCR Instrument User Manual
What is the intended use of the QuantStudio 6 Pro and 7 Pro Real-Time PCR Systems?
These systems are intended for research use only and are not for use in diagnostic procedures.
From: Applied Biosystems QuantStudio 6 Pro and 7 Pro Real-Time PCR Systems User Guide
What specific instrument functions were added in revision E.0?
Revision E.0 added protein thermal shift experiments to the instrument functions.
From: Applied Biosystems QuantStudio 6 Pro and 7 Pro Real-Time PCR Systems User Guide