Operation and maintenance manual for Tuttnauer tabletop autoclaves covering models 1730, 2340, 2540, 3140, 3850, and 3870 M & MK.
This manual provides comprehensive instructions for the operation and maintenance of Tuttnauer tabletop autoclaves, specifically models 1730, 2340, 2540, 3140, 3850, and 3870 M & MK. It covers essential topics including general information, technical data, installation procedures, preparation before sterilization, and detailed operating instructions for loading and unloading devices. The document also includes specific guidance on filling water reservoirs and sterilization time tables.
| Sterilization Temperature Range | 212ºF - 273ºF (100ºC - 134ºC) |
|---|---|
| Chamber Material | electro-polished stainless steel 316 L |
| Door Material | stainless steel CF8 |
| Maximum Altitude | 2000m |
| Minimum Room Temperature | 41ºF (5ºC) |
| Power Supply (US) | 1ph, 120V – 16A,50/60 Hz |
| Power Supply (EU) | 1ph, 230V – 16A,50/60 Hz |
| Peak Sound Level | < 70 dBA |
The autoclave is intended for use by hospital personnel and other medical personnel.
Tuttnauer warrants all new Manual autoclaves for a period of one full year, covering both parts and labor.
No, the Door Bellows is a safety device that prevents the door from opening when the chamber is pressurized.