Tuttnauer Tabletop Autoclaves Operation & Maintenance Manual
This manual provides comprehensive instructions for the operation and maintenance of Tuttnauer tabletop autoclaves, specifically models 1730, 2340, 2540, 3140, 3850, and 3870 M & MK. It covers essential topics including general information, technical data, installation procedures, preparation before sterilization, and detailed operating instructions for loading and unloading devices. The document also includes specific guidance on filling water reservoirs and sterilization time tables.
Read onlineTuttnauer 1730, 2340, 2540, 3140, 3850, 3870 M & MK specifications
| Sterilization Temperature Range | 212ºF - 273ºF (100ºC - 134ºC) |
|---|---|
| Chamber Material | electro-polished stainless steel 316 L |
| Door Material | stainless steel CF8 |
| Maximum Altitude | 2000m |
| Minimum Room Temperature | 41ºF (5ºC) |
| Power Supply (US) | 1ph, 120V – 16A,50/60 Hz |
| Power Supply (EU) | 1ph, 230V – 16A,50/60 Hz |
| Peak Sound Level | < 70 dBA |
Source: Tuttnauer Tabletop Autoclaves Operation & Maintenance Manual
All Tuttnauer 1730, 2340, 2540, 3140, 3850, 3870 M & MK documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Disassembly, diagnostics and part numbers, for repair.
Questions answered in these manuals
Who is the intended user of this autoclave?
The autoclave is intended for use by hospital personnel and other medical personnel.
From: Tuttnauer Tabletop Autoclaves Operation & Maintenance Manual
What is the warranty period for the autoclave?
Tuttnauer warrants all new Manual autoclaves for a period of one full year, covering both parts and labor.
From: Tuttnauer Tabletop Autoclaves Operation & Maintenance Manual
Who is authorized to service the autoclave?
The manual states that it is forbidden for unqualified and unauthorized personnel to service the autoclave.
From: Tuttnauer Table-Top Autoclaves Models 1730 to 3870 Technician Manual
How often should the safety valve be tested?
The safety valve must be tested once a month by operating it.
From: Tuttnauer Table-Top Autoclaves Models 1730 to 3870 Technician Manual