Tuttnauer EZ11-Plus Tabletop Autoclave Use and Maintenance Manual

Standard operating procedures for using and maintaining the Tuttnauer EZ11-Plus tabletop autoclave in a laboratory setting.


Tuttnauer EZ11-Plus Tabletop Autoclave Use and Maintenance Manual - cover page
Brand
Tuttnauer
Category
Other
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
181 KB
Published
01 October, 2026
Updated
01 October, 2026
MD5 checksum
3C6FEFCDBE3B223FE0EAF37239801EB3

About this manual

This document describes the general use, safety procedures, and maintenance tasks for the Tuttnauer EZ11-Plus tabletop autoclave installed in the UNT Lab Animal Facility. It outlines the steps for starting a sterilization cycle, loading instruments, and handling completed loads.

The SOP also details daily, weekly, and bi‑annual maintenance activities, troubleshooting guidance, and safety precautions to ensure reliable operation and compliance with laboratory standards.

What's inside

  • Purpose
  • Scope
  • General Operation of Table‑top Autoclave
  • Table‑top Autoclave Maintenance
  • Troubleshooting and Repair
  • Safety
  • References

Specifications

ModelEZ11-Plus
Serial Number18020772
Water Reservoir RequirementDistilled water only
Daily Cleaning ProcedureInspect and clean while unit is cold with mild detergent
HEPA Filter ReplacementReplace during bi‑annual maintenance

Frequently asked questions

How do you start a sterilization cycle?

Set the front‑panel switch to ON, fill the water reservoir, select the desired program, and press the START/STOP key once.

What daily maintenance is required?

While the unit is still cold, inspect and clean the exterior, door gasket, and sensors with a mild detergent and a soft, lint‑free cloth.

What should be done if sterilization indicators do not change color?

Do not consider the items sterilized; repack them and run another cycle after checking the autoclave’s operation.