Instructions for SIGNET 8510 flow transmitter, covering loop power, sensor and pulse output wiring, mounting options, functions, calibration and accessories.
The SIGNET 8510 Compak Flow Transmitter Instructions are a multilingual document in English, German and French, covering loop power connections, compatible sensor wiring, pulse output wiring and installation options. Mounting choices include standard panel mounting, the optional 3-8010 Universal Mounting Kit and the optional 3-8011 Integral Mounting Kit.
The document also provides function, view, calibrate and options examples, including K-factor calibration and display/output settings. The accessories section lists ordering codes for mounting kits and 3-8510 integral sensors.
| Loop power | 10 to 30 VDC |
|---|---|
| Loop output | 4 to 20 mA |
| Alternate loop output | 1 to 5 VDC |
| Pulse output supply | +5 to 30 VDC |
| 3-8011 compatible sensors | 3-8510-XX flow sensors only |
| NPT conduit ports | 1/2 in. NPT threaded conduit ports (2) |
| DIN conduit ports | PG13.5 threaded conduit ports (2) |
Remove power to the unit before wiring input and output connections. The warning appears in English, German and French.
The wiring examples show a 4 to 20 mA loop output.
The 3-8011 kit is compatible with 3-8510-XX flow sensors only.
Find the K-factor for your pipe size in your sensor manual.