Signet 5700 pH/ORP Monitor User Manual

This document is the user manual for the Signet 5700 pH/ORP Monitor, which introduces the product's connection, usage methods, etc.


Signet 5700 pH/ORP Monitor User Manual - cover page
Brand
Signet
Category
Health Care
Document type
User Manual
Language
English
Pages
12
File format
PDF
File size
994 KB
Published
20 October, 2023
Updated
20 October, 2023
MD5 checksum
41DABC8730BE8F1F89DA9888D9CA60F5

About this manual

This manual provides step-by-step instructions for operating and wiring the Signet 5700 pH/ORP Monitor. It covers power connections, electrode and preamp wiring, 4-20mA output, relay configurations, menu navigation, calibration procedures, and troubleshooting. The device supports both pH and ORP measurements with relay control for alarm or pump applications. Calibration is supported via EASY CAL or manual methods using standard pH buffers.

What's inside

  • Power Connections
  • Compatible Electrode/Preamp Wiring
  • 4-20mA Current Output Connections
  • Relay Connections
  • Relay Operation
  • Menu Functions
  • Calibration
  • Troubleshooting
  • Maintenance

Specifications

Power Supply12-24 VDC or 12-24 VAC
Relay Contact Ratings5 A @ 30 VDC, 5 A @ 125 VAC, or 3 A @ 250 VAC
4-20mA Output Loop Impedance350 Ω (12V supply), 950 Ω (24V supply)
Electrode Calibration BufferspH 4.0, 7.0, or 10.0

Frequently asked questions

How do I wire the 4-20mA output?

Use a 4-20mA output with maximum loop impedance of 350 Ω at 12V or 950 Ω at 24V. To isolate output and prevent ground loops, use a monitor with isolated inputs or separate DC supplies.

What are the relay operation modes?

The relay supports Low alarm, High alarm, and Proportional Pulse modes. In Proportional Pulse mode, the relay pulses to control devices like metering pumps based on pH setpoints and pulse rates.

How do I calibrate the pH using EASY CAL?

Place the electrode in a pH buffer (4.0, 7.0, or 10.0), allow stabilization, then press the first buffer calibration. For dual-point calibration, repeat with a second buffer.

What is the maximum pulse rate for relay control?

The maximum pulse rate is 300 pulses per minute, which can be set for proportional pulse mode.

Download PDF (994 KB)