Georg Fischer SIGNET 8750-1 pH/ORP Transmitter Operating Manual

This is an operating manual for the +GF+ SIGNET 8750-1 pH/ORP transmitter. It provides detailed instructions on the installation, product specifications, electrical connection, menu functions, etc.


Georg Fischer SIGNET 8750-1 pH/ORP Transmitter Operating Manual - cover page
Brand
Signet
Document type
User Manual
Language
Chinese
Pages
12
File format
PDF
File size
329 KB
Published
13 March, 2012
Updated
01 October, 2023
MD5 checksum
AED8E1811E64F2649CDF2235622099F7

About this manual

This manual covers the installation of the SIGNET 8750-1 pH/ORP transmitter, including mounting hardware, wiring connections, and power supply requirements. Detailed wiring diagrams show terminal assignments for power, sensor signals, and output loops.

It also describes the device's functional settings, calibration procedures (EASY CAL and manual), display options, and troubleshooting steps for both pH and ORP measurements, with guidance on error conditions such as out‑of‑range readings. This manual is written in Chinese.

What's inside

  • Installation
  • Product Specifications
  • Wiring Connections
  • Calibration Procedures
  • Display Settings
  • Operation Modes
  • Troubleshooting pH
  • Troubleshooting ORP

Specifications

Accuracy±0.03 mA at 25 °C, 24 V
Weight325 g (12 oz)
Operating temperature range-10 °C to 70 °C
Relative humidity range0 % to 95 %
pH measurement range0.00 – 14.00 pH
ORP measurement range-1000 mV to +2000 mV
Power supply12‑24 VDC
Ingress protectionNEMA 4, IP65
Housing materialPBT
CertificationsCSA, CE, UL

Frequently asked questions

What is the operating temperature range of the device?

The transmitter operates from –10 °C to 70 °C.

How can I troubleshoot an out‑of‑range error?

Check that the correct calibration buffers are used, verify sensor condition, and if the error persists use Manual CAL or replace the sensor.