This manual provides usage instructions, operating procedures, and maintenance guidelines for the Ohaus CN-1 composite ammonium ion electrode.
The Ohaus CN-1 is a composite ammonium ion electrode featuring a PVC membrane and a built-in reference electrode with a gel electrolyte. It is designed for the accurate and economical determination of free ammonium ions in water, fertilizers, and nitrate samples. The electrode operates within a temperature range of 0 to 50°C and has a measurement range of 1M to 5x10-6M (18,000 to 0.1 ppm as N). This manual is written in Chinese.
| Temperature Range | 0~50° C |
|---|---|
| Reference Electrolyte | Gel |
| Cable Length | 1m |
| Electrode Resistance | 1~4 MΩ |
| pH Range | 4~10 |
| Measurement Range | 1M~5x10-6M |
| Minimum Sample Volume | 50ml in beaker |
When detecting from low to high concentration, the response time is about 1 minute; from high to low, it generally takes several minutes or longer.
For direct measurement repeatability of about ±4%, the electrode should be calibrated once per hour.
The difference between two readings (E2-E1) should be 56mV ± 4mV at 25°C.
Drain the internal solution, rinse the reference pool with distilled water, install the protective cap on the sensitive part, and store it dry.