Instructions for preparing a site for the Keysight i3070 Series 5 In‑Circuit Test System, covering responsibilities, planning, and technical requirements.
This guide details the site‑preparation process for the Keysight i3070 Series 5 In‑Circuit Test System. It outlines the responsibilities of Keysight and the customer, provides a checklist, and describes the steps from planning to verification before installation.
The document includes specifications for structural, environmental, power, RF attenuation, compressed‑air, and vacuum requirements, as well as detailed receiving and moving instructions to ensure a safe and compliant installation.
| Operating Temperature | 0 °C to 40 °C (32 °F to 104 °F) |
|---|---|
| Operating Humidity | 5 % to 80 % non‑condensing |
| Storage Temperature | -40 °C to 70 °C (-40 °F to 158 °F) |
| Storage Humidity | 20 % to 80 % non‑condensing |
| Heat Dissipation (1 module) | 10 500 BTU (3.1 kW) |
| Compressed Air Max Pressure (Test head only) | 1035 kPa (150 psi) |
| Compressed Air Minimum Flow Rate (Test head only) | 0.66 l/s at STP (1.4 SCFM) |
| Compressed Air Relative Humidity Allowed | 70 % at 25 °C (150 psi) |
The system must operate in an ambient temperature range of 0 °C to 40 °C (32 °F to 104 °F).
The customer is responsible for providing insurance coverage for the system from the time it is delivered to the facility.