Keysight i3070 Series 5 In-Circuit Test System Site Preparation Guide

Instructions for preparing a site for the Keysight i3070 Series 5 In‑Circuit Test System, covering responsibilities, planning, and technical requirements.


Keysight i3070 Series 5 In-Circuit Test System Site Preparation Guide - cover page
Brand
Keysight
Category
Other Electronics
Document type
Installation Guide
Language
English
Pages
71
File format
PDF
File size
1.9 MB
Published
03 October, 2026
Updated
03 October, 2026
MD5 checksum
5A4C81B774A6432C94C0CEE24705761A

About this manual

This guide details the site‑preparation process for the Keysight i3070 Series 5 In‑Circuit Test System. It outlines the responsibilities of Keysight and the customer, provides a checklist, and describes the steps from planning to verification before installation.

The document includes specifications for structural, environmental, power, RF attenuation, compressed‑air, and vacuum requirements, as well as detailed receiving and moving instructions to ensure a safe and compliant installation.

What's inside

  • Site Preparation Process
  • Planning
  • Structural Requirements
  • RF Attenuation Requirements
  • Environmental Requirements
  • Power Requirements
  • Compressed Air and Vacuum Requirements
  • Receiving and Moving Instructions
  • Site Preparation Checklist

Specifications

Operating Temperature0 °C to 40 °C (32 °F to 104 °F)
Operating Humidity5 % to 80 % non‑condensing
Storage Temperature-40 °C to 70 °C (-40 °F to 158 °F)
Storage Humidity20 % to 80 % non‑condensing
Heat Dissipation (1 module)10 500 BTU (3.1 kW)
Compressed Air Max Pressure (Test head only)1035 kPa (150 psi)
Compressed Air Minimum Flow Rate (Test head only)0.66 l/s at STP (1.4 SCFM)
Compressed Air Relative Humidity Allowed70 % at 25 °C (150 psi)

Frequently asked questions

What temperature range is required for operating the i3070 system?

The system must operate in an ambient temperature range of 0 °C to 40 °C (32 °F to 104 °F).

Who must provide insurance for the i3070 system?

The customer is responsible for providing insurance coverage for the system from the time it is delivered to the facility.

Download PDF (1.9 MB)