Instructions for preparing a site for the Keysight i3070 Series 5 In‑Circuit Test System, covering responsibilities, planning, and technical requirements.
About this manual
This guide details the site‑preparation process for the Keysight i3070 Series 5 In‑Circuit Test System. It outlines the responsibilities of Keysight and the customer, provides a checklist, and describes the steps from planning to verification before installation.
The document includes specifications for structural, environmental, power, RF attenuation, compressed‑air, and vacuum requirements, as well as detailed receiving and moving instructions to ensure a safe and compliant installation.
What's inside
- Site Preparation Process
- Planning
- Structural Requirements
- RF Attenuation Requirements
- Environmental Requirements
- Power Requirements
- Compressed Air and Vacuum Requirements
- Receiving and Moving Instructions
- Site Preparation Checklist
Specifications
| Operating Temperature | 0 °C to 40 °C (32 °F to 104 °F) |
|---|
| Operating Humidity | 5 % to 80 % non‑condensing |
|---|
| Storage Temperature | -40 °C to 70 °C (-40 °F to 158 °F) |
|---|
| Storage Humidity | 20 % to 80 % non‑condensing |
|---|
| Heat Dissipation (1 module) | 10 500 BTU (3.1 kW) |
|---|
| Compressed Air Max Pressure (Test head only) | 1035 kPa (150 psi) |
|---|
| Compressed Air Minimum Flow Rate (Test head only) | 0.66 l/s at STP (1.4 SCFM) |
|---|
| Compressed Air Relative Humidity Allowed | 70 % at 25 °C (150 psi) |
Frequently asked questions
What temperature range is required for operating the i3070 system?
The system must operate in an ambient temperature range of 0 °C to 40 °C (32 °F to 104 °F).
Who must provide insurance for the i3070 system?
The customer is responsible for providing insurance coverage for the system from the time it is delivered to the facility.