Hioki 8938 FFT Analog Unit Instruction Manual

Hioki 8938 FFT Analog Unit instruction manual covering safety precautions, usage notes, specifications and the module replacement procedure.


Hioki 8938 FFT Analog Unit Instruction Manual - cover page
Brand
Hioki
Category
Other Electronics
Document type
User Manual
Language
English
Pages
1
File format
PDF
File size
154 KB
Published
07 November, 2026
Updated
07 November, 2026
MD5 checksum
14E55A8746C8147F2D619FDDC02459D0

About this manual

This is the instruction manual for the Hioki Model 8938 FFT Analog Unit, an FFT analog unit for Hioki Memory HiCorders. It lists the Hioki devices the unit is supported in (8720, 8826, 8835, 8835-01, 8841, 8842, 8860 and 8861), explains initial inspection and preliminary checks, and gives safety precautions with the CAT I to CAT IV measurement category definitions.

The manual also contains usage notes, maintenance and service instructions, a full specifications table for the 8938 (input channels, measurement ranges, accuracy, frequency characteristic, filters, input ratings, dimensions and mass), and a step-by-step procedure for removing and replacing the unit in the main instrument.

What's inside

  • Introduction
  • Initial Inspection
  • Preliminary Checks
  • Maintenance and Service
  • Safety
  • Safety Symbol
  • Measurement categories (Overvoltage categories)
  • Usage Notes
  • Specifications
  • Replacement Procedure

Specifications

Number of input channels2
A/D resolution12 bits
Maximum sampling rate1 MS/s (sampling period: 1 μs)
Input typeUnbalanced (input isolated from output)
Input resistance1 MΩ ±1%
Maximum input voltage400 V DC
Maximum rated voltage to earth370 V DC
Frequency characteristicDC to 400 kHz ±3 dB (DC coupling); 7 Hz to 400 kHz ±3 dB (AC coupling)
Common mode rejection ratio80 dB minimum (at 50/60 Hz, signal source resistance 100 Ω maximum)
Dimensions and massApprox. 170W x 20H x 148.5D mm (excluding projections), approx. 290 g
Storage temperature and humidity-10 to 50°C (14 to 122°F), 80% RH or less (no condensation)

Frequently asked questions

Which Hioki devices support the 8938 FFT Analog Unit?

The manual lists 8720, 8826 (ROM Ver.2.00 or later), 8835 (ROM Ver.5.00 or later), 8835-01, 8841, 8842, 8860 and 8861 as supported Hioki devices.

What is the maximum input voltage of the 8938?

The maximum input voltage is 400 V DC. The maximum rated voltage between the input terminals and ground is 370 V DC.

Which connection cords should be used with the unit?

For safety reasons, only the specified Hioki 9197/9198 Connection Cord options should be used when taking measurements.

How do I remove the 8938 from the main unit?

Remove the connection cords, power off the main unit and disconnect the power cord, take out the two fixing screws with a Phillips screwdriver, then grasp the handle and pull the module out.