User manual for Hanna HI 9147 portable dissolved oxygen meter with galvanic probe, covering warranty, setup, calibration, and maintenance.
This manual provides instructions for using the Hanna HI 9147 portable dissolved oxygen meter. It includes warranty details, initial inspection, product configuration, calibration, and maintenance procedures. The device measures dissolved oxygen in mg/L or % saturation, with automatic temperature compensation and salinity adjustment. It is designed for field and lab use in water quality testing.
The meter comes with a fixed galvanic probe (HI 76409/4 or HI 76409/10), spare membranes, electrolyte solution, and batteries. Calibration is manual and performed in air. Temperature and salinity compensation are automatic. The device is waterproof and suitable for harsh environments with optional protective sleeves. This manual is written in Spanish.
| Range | 0.0 to 50.0 mg/L (ppm) O2 or 0 to 600 % O2 |
|---|---|
| Temperature Range | -5.0 to 50.0°C |
| Resolution | 0.1 mg/L or 1% (O2) / 0.1°C (1°F) |
| Accuracy | ±1% of reading (O2) / ±0.2°C (1°F) (excluding probe error) |
| Salinity Compensation | 0 to 51 g/L (resolution 1 g/L) |
| Probe | HI 76409/4 with 4m cable or HI 76409/10 with 10m cable |
| Battery Type | 3 x 1.5V AAA |
| Battery Life | 1000 hours (BL off) / 5 years |
| Dimensions | 185 x 72 x 36 mm (7.3 x 2.8 x 1.4") |
| Weight | 450 g (15.9 oz) with HI 76409/4 probe |
Calibration is a one-point procedure performed in air. Ensure the probe is prepared (membrane filled with electrolyte) and turn on the meter. Select the calibration mode and set to 100%.
The temperature range is -5.0 to 50.0°C (-5.0 to 122.0°F).
The membrane should be replaced every 2 months for optimal performance.
The meter is guaranteed for 2 years against manufacturing and material defects when used for intended purposes and maintained according to instructions. Probes are guaranteed for 1 year.