Instruction manual for the Hanna HI 3812 Hardness Test Kit, covering EDTA titration procedure, ranges, specifications, and kit contents.
The Hanna HI 3812 Hardness Test Kit determines water hardness as mg/L (ppm) calcium carbonate by EDTA titration. The kit includes Hardness Buffer, Calmagite Indicator, HI 3812-0 EDTA Solution, plastic beakers, a syringe, and a tip, with instructions for high and low range testing.
The high range procedure uses a 5 mL sample and multiplies the titrant volume by 300; the low range procedure uses a 50 mL sample and multiplies by 30 for improved precision below 30 mg/L.
| Range | 0.0 to 30.0 mg/L (ppm) CaCO3; 0 to 300 mg/L (ppm) CaCO3 |
|---|---|
| Smallest Increment | 0.3 mg/L in the 0-30 mg/L range; 3 mg/L in the 0-300 mg/L range |
| Analysis Method | EDTA titration |
| Sample Size | 5 mL or 50 mL |
| Number of Tests | 100 (average) |
| Case Dimensions | 200x120x60 mm (7.9x4.7x2.4 in) |
| Shipping Weight | 460 g (1 lb.) |
It uses EDTA titration to determine hardness as mg/L (ppm) calcium carbonate.
The kit provides 100 tests on average.
The procedure uses either a 5 mL or a 50 mL sample.
Read the titrant volume in milliliters and multiply by 300 to obtain mg/L (ppm) CaCO3.