Hanna HI 9143, HI 9143M, HI 9145 Dissolved Oxygen Meter Instruction Manual

Instruction manual for Hanna HI 9143, HI 9143M, and HI 9145 portable dissolved oxygen meters, covering calibration, measurement, and maintenance.


Hanna HI 9143, HI 9143M, HI 9145 Dissolved Oxygen Meter Instruction Manual - cover page
Brand
Hanna
Category
Analyzer
Document type
User Manual
Language
English
Pages
20
File format
PDF
File size
113 KB
Published
06 November, 2026
Updated
06 November, 2026
MD5 checksum
17D1DDEA61964549EB1A58D6340FD813

About this manual

The Hanna HI 9143 and HI 9145 are water-resistant, microprocessor-based, auto-calibration dissolved oxygen meters with automatic temperature compensation. They are described for dissolved oxygen and temperature measurement in water and wastewater, as well as other applications such as fish farming.

The meter displays dissolved oxygen in ppm or % saturation and temperature from 0 to 50°C. The manual covers preliminary examination, probe and meter descriptions, specifications, probe preparation, calibration, measurement, altitude and salinity compensation, probe and membrane maintenance, battery replacement, accessories, and warranty.

What's inside

  • Preliminary Examination
  • General Description
  • Functional Description - Probe
  • Functional Description - Meter
  • Specifications
  • Probe Preparation
  • Calibration
  • Measurement
  • Altitude Compensation
  • Salinity Compensation

Specifications

Range0.00 to 45.00 mg/L O2; 0.0 to 300 %O2; 0.0 to 50.0 °C
Resolution0.01 mg/L O2; 0.1 %O2; 0.1 °C
Accuracy±1.5% Full Scale mg/L O2; ±1.5% Full Scale %O2; ±0.5 °C
CalibrationAutomatic in saturated air
Temperature compensationAutomatic from 0 to 50°C (32 to 122°F)
Altitude compensation0 to 1900 m, 100 m resolution
Salinity compensation0 to 40 g/L, 1 g/L resolution
Battery4 x 1.5V AA, 200 hours continuous use; auto shut-off after 4 hours
Dimensions196 x 80 x 60 mm (7.7 x 3.1 x 2.4 in)
Shipping weight425 g (15 oz.); kit 1.4 kg (3.1 lb.)

Frequently asked questions

How often should the dissolved oxygen meter be calibrated?

Calibration is recommended every time the meter is switched on. The instrument must also be calibrated whenever the probe, membrane or electrolyte is changed.

What water movement is required for accurate dissolved oxygen measurements?

A water movement of at least 30 cm (12”) per second is required to keep oxygen replenished at the membrane surface.

What does COND on the display mean?

COND indicates that the probe is in auto-conditioning (automatic polarization) mode. Once it disappears, the probe is polarized and the instrument can be calibrated.

How long is the probe warranty?

The probes are warranted for a period of six months.

Download PDF (113 KB)

More Hanna Manuals