Instruction manual for Hanna HI 9143, HI 9143M, and HI 9145 portable dissolved oxygen meters, covering calibration, measurement, and maintenance.
The Hanna HI 9143 and HI 9145 are water-resistant, microprocessor-based, auto-calibration dissolved oxygen meters with automatic temperature compensation. They are described for dissolved oxygen and temperature measurement in water and wastewater, as well as other applications such as fish farming.
The meter displays dissolved oxygen in ppm or % saturation and temperature from 0 to 50°C. The manual covers preliminary examination, probe and meter descriptions, specifications, probe preparation, calibration, measurement, altitude and salinity compensation, probe and membrane maintenance, battery replacement, accessories, and warranty.
| Range | 0.00 to 45.00 mg/L O2; 0.0 to 300 %O2; 0.0 to 50.0 °C |
|---|---|
| Resolution | 0.01 mg/L O2; 0.1 %O2; 0.1 °C |
| Accuracy | ±1.5% Full Scale mg/L O2; ±1.5% Full Scale %O2; ±0.5 °C |
| Calibration | Automatic in saturated air |
| Temperature compensation | Automatic from 0 to 50°C (32 to 122°F) |
| Altitude compensation | 0 to 1900 m, 100 m resolution |
| Salinity compensation | 0 to 40 g/L, 1 g/L resolution |
| Battery | 4 x 1.5V AA, 200 hours continuous use; auto shut-off after 4 hours |
| Dimensions | 196 x 80 x 60 mm (7.7 x 3.1 x 2.4 in) |
| Shipping weight | 425 g (15 oz.); kit 1.4 kg (3.1 lb.) |
Calibration is recommended every time the meter is switched on. The instrument must also be calibrated whenever the probe, membrane or electrolyte is changed.
A water movement of at least 30 cm (12”) per second is required to keep oxygen replenished at the membrane surface.
COND indicates that the probe is in auto-conditioning (automatic polarization) mode. Once it disappears, the probe is polarized and the instrument can be calibrated.
The probes are warranted for a period of six months.