Hanna pH300/pH301/pH302 Microprocessor Bench-top pH/mV/ISE/°C Meters Instruction Manual

Instruction manual for Hanna pH300, pH301 and pH302 microprocessor bench-top meters, covering setup, operation, calibration, logging, programming and…


Hanna pH300/pH301/pH302 Microprocessor Bench-top pH/mV/ISE/°C Meters Instruction Manual - cover page
Brand
Hanna
Category
Analyzer
Document type
User Manual
Language
English
Pages
57
File format
PDF
File size
859 KB
Published
08 November, 2026
Updated
08 November, 2026
MD5 checksum
1DC26ED1D92DD9FD1CE0133186F4F7DB

About this manual

The Hanna pH300, pH301 and pH302 are microprocessor-controlled bench-top meters for pH, ORP, ISE and temperature measurements. They provide automatic temperature compensation and an LCD that shows measurement and temperature information.

The instruction manual covers front and rear panel descriptions, specifications for each model, operational guidance, pH/mV/ION/temperature calibration, logging with pH301, programming functions with pH302, PC interface details, troubleshooting and electrode maintenance.

What's inside

  • General Description
  • Functional Description pH300
  • Specifications pH300
  • Specifications pH301
  • Specifications pH302
  • pH Calibration
  • ION Calibration (for pH301 only)
  • Logging with pH301
  • Programming Functions with pH302
  • Troubleshooting Guide

Specifications

pH range (pH300/pH302)0.00 to 14.00
pH range (pH301)-1.999 to 19.999
pH resolution (pH300/pH302)0.01
pH accuracy (pH300/pH302)±0.01 at 20°C/68°F
Temperature range-9.9 to 120.0 °C
Power supply12 VDC
Computer interfaceRS 232, optoisolated
Logging interval (pH301)1, 15, 30 seconds or 1, 5, 30, 60, 120, 180 minutes
Printer (pH302)Dot matrix, 44 mm wide paper

Frequently asked questions

What is included with each pH meter?

Each meter comes with an HI 1131B pH electrode, HI 7669/2W temperature probe, HI 3131B ORP electrode, pH 4.01 and 7.01 buffer solutions and a 12 VDC adapter; pH302 also includes 5 paper rolls.

How many pH calibration points does pH301 support?

pH301 supports automatic 1, 2 or 3 point pH calibration with six memorized buffers.

What interface is available for connecting to a PC?

The meters have an optoisolated RS 232 interface for transferring sample measurements to a PC.

Can pH301 store measurements over time?

Yes. pH301 has a selectable logging interval of 1, 15, 30 seconds or 1, 5, 30, 60, 120, 180 minutes.

Download PDF (859 KB)