The HI 3811 Alkalinity Test Kit is a chemical test kit that is used to measure the alkalinity of water. The kit uses a titration method, using phenolphthalein and bromophenol blue indicators, to measure alkalinity. The kit has a range of 0 to 100 mg/l (ppm) CaCO3, with an increment of 1 mg/l.
This is the Spanish instruction manual (ISTR3811R1) for the Hanna HI 3811 Alkalinity Test Kit. It explains that alkalinity is the quantitative capacity of a water sample to neutralize an acid to a set pH, which matters for assessing the corrosive capacity of water from hydroxide, carbonate and bicarbonate ions, and for drinking water, wastewater, boiler and cooling system treatment and soils.
The manual gives the kit specifications (ranges of 0 to 100 and 0 to 300 mg/l CaCO3, titration method with phenolphthalein and bromophenol blue), lists the supplied reagents and calibrated vessels, and provides step-by-step instructions for determining phenolphthalein alkalinity, total alkalinity and the low-range procedure, plus references and health and safety notes. This manual is written in Spanish.
| Range | 0 to 100 mg/l (ppm) CaCO3; 0 to 300 mg/l (ppm) CaCO3 |
|---|---|
| Smallest increment | 1 mg/l (0-100 mg/L range); 3 mg/l (0-300 mg/l range) |
| Analysis method | Acid titration using phenolphthalein and bromophenol blue |
| Sample size | 5 ml and 15 ml |
| Number of tests | 110 (average) |
| Box dimensions | 200x120x60 mm |
| Weight | 460 g |
Each kit is supplied with Reactant 1 (10 ml dropper bottle), Reactant 2 (10 ml dropper bottle), Reactant 3 (120 ml bottle), two calibrated vessels (10 and 50 ml) and one calibrated syringe.
Read the milliliters on the syringe scale and multiply by 300 for the 5 ml sample; for the 15 ml low-range sample, multiply the syringe reading by 100.
Fill the calibrated vessel to the 15 ml mark instead of 5 ml and follow the same test procedure, then multiply the syringe reading by 100.