Quick reference guide for the Hamilton Medical HAMILTON-MR1 MRI-compatible ventilator covering setup, operation, and MR environment safety.
This Quick Guide provides a concise reference for ventilating adult and pediatric patients with the HAMILTON-MR1 ventilator. It covers the device basics, touchscreen navigation, breathing circuit and nebulizer connections, power-up sequence, and sensor monitoring. The guide emphasizes safe operation in the magnetic resonance (MR) environment, including TeslaSpy magnetic field monitoring and positioning rules.
Additional sections detail patient configuration, preoperational checks, ventilation mode settings, CPR ventilation, patient monitoring, ventilation modes, monitoring parameters, and control settings. The document does not replace the full Operator's Manual and notes that some functions are optional and market-dependent.
| Software version | v3.0.x (2020-09-15) |
|---|---|
| TeslaSpy sampling rate | Once every 2 seconds |
| Maximum movement speed in MR environment | 10 cm/s (3 in/s) |
| Oxygen inlet fitting | High-pressure oxygen DISS or NIST |
| O2 sensor default | Enabled by default |
The magnetic field is within acceptable limits and the device is in a safe position.
Do not move the ventilator faster than 10 cm (3 in) per second.
Immediately move the ventilator away from the MRI scanner, provide alternative ventilation, disconnect the patient, remove the ventilator from use, and have it serviced.