Operation manual for the Hamilton Medical VENTILAIRII medical air compressor, covering setup, use, maintenance and safety.
The VENTILAIRII medical air compressor provides medical‑grade compressed air for Hamilton Medical ventilators, serving as a primary or backup air source in hospitals. The manual describes the device’s physical features, electrical connections, and operating procedures, and includes safety warnings and regulatory information.
Only trained personnel should operate or service the unit. The guide details preventive maintenance, troubleshooting steps, replacement parts, and technical specifications such as voltage options, pressure limits, and fuse requirements. This manual is written in Chinese.
| Input voltage (Version 1) | 100 to 115 V ±10%, 50/60 Hz (PN 155601) |
|---|---|
| Input voltage (Version 2) | 220 to 240 V ±10%,50 Hz / 230 V ±10%,60 Hz (PN 155600) |
| Fuse rating (100‑115 V version) | T 10 A H 250 V |
| Fuse rating (220‑240 V version) | T 3.15 A H 250 V |
| Air inlet pressure range | 2 – 6 bar (peak up to 10 bar) |
| Backup de‑activation pressure | Above 260 kPa (40 psi, 2.6 bar) |
Two versions are offered: 100‑115 V ±10 % at 50/60 Hz (PN 155601) and 220‑240 V ±10 % at 50 Hz / 230 V ±10 % at 60 Hz (PN 155600).
Use a T 3.15 A H 250 V fuse for the 220‑240 V model.
No, the compressor is an oil‑free design and must not be lubricated with any oil.