This manual details the ventilation modes, technical specifications, control settings, and monitoring parameters for the Hamilton Medical MR1 ventilator.
The Hamilton Medical MR1 is a ventilator designed for Adult/Pediatric and Neonatal patients. It supports a wide range of ventilation modes, including volume-targeted modes (APVcmv, APVsimv), pressure-controlled modes (PCV+, PSIMV+, DuoPAP), and noninvasive modes (NIV, nCPAP). The device also features intelligent ventilation (ASV) and high-flow oxygen therapy (HiFlowO2).
| Weight | 6.8 kg (15 lb) |
|---|---|
| Weight with trolley | 21 kg (46.2 lb) |
| Maximum inspiratory flow | 260 l/min |
| Inspiratory pressure range | 0 to 60 cmH2O |
| Tidal volume (Adult/Ped) | 20 to 2000 ml |
| Tidal volume (Neonatal) | 2 to 300 ml |
| Input power | 100 to 240 VAC ±10%, 50/60 Hz |
| Battery operating time (80% brightness) | 8 h |
| Battery operating time (20% brightness) | 9.2 h |
The PEEP setting range for Neonatal patients is 3 to 25 cmH2O.
Yes, the device is MR Conditional up to 50 mT.