Hamilton Medical MR1 Ventilator User Manual

This manual details the ventilation modes, technical specifications, control settings, and monitoring parameters for the Hamilton Medical MR1 ventilator.


Hamilton Medical MR1 Ventilator User Manual - cover page
Brand
Hamilton Medical
Category
Other
Document type
User Manual
Language
English
Pages
8
File format
PDF
File size
325 KB
Published
29 October, 2026
Updated
29 October, 2026
MD5 checksum
10DB3BBEAD6134E3F9265BC785EE7DB2

About this manual

The Hamilton Medical MR1 is a ventilator designed for Adult/Pediatric and Neonatal patients. It supports a wide range of ventilation modes, including volume-targeted modes (APVcmv, APVsimv), pressure-controlled modes (PCV+, PSIMV+, DuoPAP), and noninvasive modes (NIV, nCPAP). The device also features intelligent ventilation (ASV) and high-flow oxygen therapy (HiFlowO2).

What's inside

  • Ventilation modes
  • Standard configuration and options
  • Technical performance
  • Pneumatic performance
  • Electrical specifications
  • Battery specifications
  • Graphical patient data
  • Alarms
  • Control settings and ranges
  • Monitoring parameters

Specifications

Weight6.8 kg (15 lb)
Weight with trolley21 kg (46.2 lb)
Maximum inspiratory flow260 l/min
Inspiratory pressure range0 to 60 cmH2O
Tidal volume (Adult/Ped)20 to 2000 ml
Tidal volume (Neonatal)2 to 300 ml
Input power100 to 240 VAC ±10%, 50/60 Hz
Battery operating time (80% brightness)8 h
Battery operating time (20% brightness)9.2 h

Frequently asked questions

What is the range for the PEEP setting in Neonatal mode?

The PEEP setting range for Neonatal patients is 3 to 25 cmH2O.

Does the MR1 support MRI compatibility?

Yes, the device is MR Conditional up to 50 mT.