Hamilton Medical ASV1 ASV Quick Guide covers Adaptive Support Ventilation basics, preparation, settings, graphs, monitoring, adjustment, and weaning.
The Hamilton Medical ASV1 ASV Quick Guide is a short clinical guide to Adaptive Support Ventilation (ASV). It covers ASV basics, preparing for ventilation, initial settings, the ASV graph, working principles, monitoring, adjustment, and weaning.
The guide includes an appendix on lung-protective rules, the ASV safety box, and a glossary. A warning states that the guide is based on clinician evaluations within and outside Hamilton Medical, is intended as an example, and does not replace the ventilator operator’s manual or a physician’s clinical judgment.
| Suggested initial %MinVol | 100% for a normal patient; 120% for ARDS |
|---|---|
| Adult minute volume calculation | 0.1 l per kg of ideal body weight (IBW) |
| Suggested initial PEEP/CPAP | 5 cmH2O |
| Suggested initial oxygen | 50% |
| Default Pasvlimit for normal patients | 30 cmH2O |
It suggests 100% for a normal patient and 120% for ARDS.
The guide suggests 5 cmH2O, or according to your ICU standard.
For adults, minute volume is calculated at 0.1 l per kg of ideal body weight (IBW).
No. It says it does not replace the official operator’s manual or the clinical judgment of a physician.