Hamilton Medical HAMILTON-C2 Ventilator Operator's Manual Addendum

Hamilton Medical AG provides online training for users via the Hamilton Medical College. See https://www.hamilton- medical.com/E-Learning-and-Education/ College.html.


Hamilton Medical HAMILTON-C2 Ventilator Operator's Manual Addendum - cover page
Brand
Hamilton Medical
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
28
File format
PDF
File size
433 KB
Published
24 November, 2023
Updated
24 November, 2023
MD5 checksum
ED14E05FB1CFD4D540EA2BC84A1C65FD

About this manual

Hamilton Medical HAMILTON-C2 ventilator addendum summarizes changes and corrections to the operator's manual, organized by affected manual sections. It includes updated safety messages, trolley and intrahospital transport requirements, alarm troubleshooting, suctioning guidance, and corrected physical, electrical, and monitored-parameter specifications.

The document also lists corrected trolley and accessory part numbers and notes online training through Hamilton Medical College. It is intended to be stored with the ventilator operator's manual.

What's inside

  • Preface: Training
  • Safety message updates
  • Section 1.3.1 Breathing circuits and accessories
  • Section 2.11 Working with the trolley
  • Section 8.5 Alarm troubleshooting table
  • Section 9.5 Suctioning maneuver
  • Section A.1 Physical characteristics
  • Section A.4 Electrical specifications
  • Section A.6 Monitored parameters
  • Chapter G Parts and Accessories

Specifications

Ventilation unit weight9.5 kg (21 lb)
Ventilation unit with standard trolley weight37 kg (81.6 lb)
Trolley maximum safe working load80 kg (176 lb)
Battery14.4 V DC, 6.8 Ah, 98 Wh; Lithium-ion
Battery recharge timeApproximately 2.45 h for one battery; approximately 5 h for two batteries
Battery storage temperature-20°C to 50°C
Pinsp range0 to 50 cmH2O
PetCO2 range0 to 150 mmHg
Vt/IBW range2 to 20 ml/kg
Vt/Weight range2 to 20 ml/kg

Frequently asked questions

Where can users find online training for the HAMILTON-C2?

Hamilton Medical AG provides online training for users via the Hamilton Medical College.

Which components may be connected during intrahospital transport?

Only the breathing circuit, flow sensor, CO2 sensor, O2 cylinder, Aerogen nebulizer, humidifier, water bottle, and infusion arm are allowed.

How long does it take to recharge the ventilator battery?

Approximately 2.45 hours for one battery and approximately 5 hours for two batteries.

What is the minimum inspiratory pressure in APVcmv and APVsimv modes?

The minimum inspiratory pressure (Ppeak - PEEP) is 5 cmH2O.