Addendum to the HAMILTON-C1/T1/MR1 ventilator operator's manuals covering corrections, alarm troubleshooting, Bluetooth setup, water protection, and updated…
Store this addendum with the HAMILTON-C1/T1/MR1 ventilator Operator's Manuals; it summarizes changes and corrections to those manuals. It covers use of the COM1 RS-232 port with the communication Y-cable, an alarm not described in the operator's manual, closed-suctioning maneuvers, Bluetooth connection setup, and Vt/Weight shown as Vt/kg in the Controls window for neonatal modes.
Additional corrections cover HAMILTON-C1/T1 water protection ratings, HAMILTON-T1 storage temperature when not in its original packaging, and updated HAMILTON-T1 standards and approvals. The document also addresses Pulse Oximetry operating conditions and INTELLiVENT-ASV Vent Status, Quick Wean, and waveform view 7.
| HAMILTON-C1 ingress protection | IP22; protected against dripping water when tilted to a maximum of 15 degrees and from solid particles larger than 12.5 mm |
|---|---|
| HAMILTON-T1 ingress protection | IP54; protected from water spray from any direction and from limited dust ingress |
| HAMILTON-T1 storage temperature without original packaging | -15°C to 60°C |
| Device classification | Class II, Type BF applied part, continuous operation |
| COM1 communication cable | Communication Y-cable, PN 10077038 |
The COM1 RS-232 connection requires the communication Y-cable, part number 10077038.
Check patient condition and the CO2 sensor and adapter connections; ensure airway adapters are not horizontal relative to the floor.
No. Ventilation continues and the current settings do not need to be adjusted.
The permitted storage temperature range is -15°C to 60°C.