Hamilton Medical HAMILTON-C1/T1/MR1 Ventilator Operator's Manuals Addendum

Addendum to the HAMILTON-C1/T1/MR1 ventilator operator's manuals covering corrections, alarm troubleshooting, Bluetooth setup, water protection, and updated…


Hamilton Medical HAMILTON-C1/T1/MR1 Ventilator Operator's Manuals Addendum - cover page
Brand
Hamilton Medical
Category
Professional medical equipment
Document type
Other
Language
English
Pages
40
File format
PDF
File size
768 KB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
9BAD6731303FCEDE189D70FE38C6C8B2

About this manual

Store this addendum with the HAMILTON-C1/T1/MR1 ventilator Operator's Manuals; it summarizes changes and corrections to those manuals. It covers use of the COM1 RS-232 port with the communication Y-cable, an alarm not described in the operator's manual, closed-suctioning maneuvers, Bluetooth connection setup, and Vt/Weight shown as Vt/kg in the Controls window for neonatal modes.

Additional corrections cover HAMILTON-C1/T1 water protection ratings, HAMILTON-T1 storage temperature when not in its original packaging, and updated HAMILTON-T1 standards and approvals. The document also addresses Pulse Oximetry operating conditions and INTELLiVENT-ASV Vent Status, Quick Wean, and waveform view 7.

What's inside

  • Corrections to the HAMILTON-C1/T1/MR1 Operator's Manuals
  • Using the COM1 RS-232 port
  • Alarm troubleshooting
  • Closed-suctioning maneuvers
  • Setting up a Bluetooth connection
  • Vt/Weight in Controls window
  • HAMILTON-C1/T1 water protection
  • HAMILTON-T1 environmental requirements
  • HAMILTON-T1 standards and approvals
  • Waveforms, view 7

Specifications

HAMILTON-C1 ingress protectionIP22; protected against dripping water when tilted to a maximum of 15 degrees and from solid particles larger than 12.5 mm
HAMILTON-T1 ingress protectionIP54; protected from water spray from any direction and from limited dust ingress
HAMILTON-T1 storage temperature without original packaging-15°C to 60°C
Device classificationClass II, Type BF applied part, continuous operation
COM1 communication cableCommunication Y-cable, PN 10077038

Frequently asked questions

What cable is required to use the COM1 RS-232 port on the HAMILTON-C1/T1?

The COM1 RS-232 connection requires the communication Y-cable, part number 10077038.

What should be checked when the CO2: Poor signal alarm occurs?

Check patient condition and the CO2 sensor and adapter connections; ensure airway adapters are not horizontal relative to the floor.

Do ventilator settings need to be adjusted during closed suctioning?

No. Ventilation continues and the current settings do not need to be adjusted.

What is the permitted storage temperature range for the HAMILTON-T1 without its original packaging?

The permitted storage temperature range is -15°C to 60°C.

Download PDF (768 KB)

More Hamilton Medical Manuals


View all Hamilton Medical manuals →