This document describes the reprocessing guide for Hamilton Medical's Autoclavable Expiratory Valve Set.
This guide explains how to reprocess the Hamilton Medical Autoclavable Expiratory Valve Set for reuse. It covers safety information, the components and materials of the valve sets used with HAMILTON-C1/T1/MR1, HAMILTON-C2/C3/C6, and HAMILTON-G5/S1 ventilators, and the reprocessing steps of disassembly, cleaning, visual inspection, packaging, sterilization, and storage.
The guide provides manual and mechanical cleaning methods, recommended cleaning agents and equipment, program parameters, and instructions for using a reprocessed valve set, disposal, and parts and accessories.
| Material heat resistance | Up to 140°C (284°F) |
|---|---|
| Manual cleaning soak time | 30 minutes |
| Manual cleaning soak temperature | 40°C (104°F) |
| Mechanical cleaning cycle | 55°C (131°F) for 5 min |
| Thermal disinfection cycle | 83°C (181.4°F) for 10 min |
| Drying cycle | 100°C (212°F) for 10 min and 95°C (203°F) for 30 min |
| Packaging heat resistance | Up to 141°C (285.8°F) |
| Reprocessing standard | EN ISO 17664 |
Reprocess the expiratory valve set immediately after use. The process includes disassembly, cleaning, a visual test, packaging, sterilization, and storage for future use.
No. Rinse aids can cause premature damage, reduce product life span, and lead to early failure and cracks in the plastic.
The valve set must be macroscopically clean and free of visible residual matter and other impurities. Check for cracks, broken or deformed parts, or discoloration, and repeat cleaning if impurities remain.
Use packaging that conforms to ISO 11607, is suitable for vapor sterilization, and is sufficiently permeable to vapor. Hamilton Medical recommends double packaging.