Hamilton Medical Autoclavable Expiratory Valve Set Reprocessing Guide

This document describes the reprocessing guide for Hamilton Medical's Autoclavable Expiratory Valve Set.


Hamilton Medical Autoclavable Expiratory Valve Set Reprocessing Guide - cover page
Brand
Hamilton Medical
Category
Home medical equipment
Document type
User Manual
Language
English
Pages
66
File format
PDF
File size
4.3 MB
Published
24 November, 2023
Updated
24 November, 2023
MD5 checksum
2B7E1FBFFD36FF7FB8095D76EF80BFEA

About this manual

This guide explains how to reprocess the Hamilton Medical Autoclavable Expiratory Valve Set for reuse. It covers safety information, the components and materials of the valve sets used with HAMILTON-C1/T1/MR1, HAMILTON-C2/C3/C6, and HAMILTON-G5/S1 ventilators, and the reprocessing steps of disassembly, cleaning, visual inspection, packaging, sterilization, and storage.

The guide provides manual and mechanical cleaning methods, recommended cleaning agents and equipment, program parameters, and instructions for using a reprocessed valve set, disposal, and parts and accessories.

What's inside

  • Safety Information
  • Overview
  • Components and materials
  • Disassembly
  • Cleaning
  • Visual test
  • Packaging
  • Sterilize
  • Store for future use
  • Disposal

Specifications

Material heat resistanceUp to 140°C (284°F)
Manual cleaning soak time30 minutes
Manual cleaning soak temperature40°C (104°F)
Mechanical cleaning cycle55°C (131°F) for 5 min
Thermal disinfection cycle83°C (181.4°F) for 10 min
Drying cycle100°C (212°F) for 10 min and 95°C (203°F) for 30 min
Packaging heat resistanceUp to 141°C (285.8°F)
Reprocessing standardEN ISO 17664

Frequently asked questions

When should the expiratory valve set be reprocessed?

Reprocess the expiratory valve set immediately after use. The process includes disassembly, cleaning, a visual test, packaging, sterilization, and storage for future use.

Can rinse aids be used when cleaning the expiratory valve set?

No. Rinse aids can cause premature damage, reduce product life span, and lead to early failure and cracks in the plastic.

What should be checked after each cleaning cycle?

The valve set must be macroscopically clean and free of visible residual matter and other impurities. Check for cracks, broken or deformed parts, or discoloration, and repeat cleaning if impurities remain.

What packaging is recommended for vapor sterilization?

Use packaging that conforms to ISO 11607, is suitable for vapor sterilization, and is sufficiently permeable to vapor. Hamilton Medical recommends double packaging.