GenScript SurePage Gels 10×8 Protein Electrophoresis User Guide

GenScript guide for using SurePage Gels 10×8 in protein electrophoresis, covering compatible gel tanks, sample prep, loading, and running conditions.


GenScript SurePage Gels 10×8 Protein Electrophoresis User Guide - cover page
Brand
GenScript
Category
medical technology
Document type
User Manual
Language
English
Pages
6
File format
PDF
File size
1.4 MB
Published
10 November, 2026
Updated
10 November, 2026
MD5 checksum
1C6F66AA8EA5427C054031134E05FE80

About this manual

This guide explains how to use SurePage™ Gels, 10×8, precast gels for protein electrophoresis from GenScript. It lists compatible gel tanks, including the Bio-Rad Mini-Protean (II/3/Tetra System), Hoefer Mighty Small (SE 250/SE 260), and Invitrogen Novex XCell I, II & SureLock, and covers both MOPS and MES buffer systems.

The numbered steps cover choosing a gel percentage, preparing samples with 4×LDS sample buffer, preparing the gel tank, peeling the tape and removing the comb, loading samples, running conditions at 200 V, and opening the cassette to remove and stain the gel. GenScript support contact details are included at the end.

What's inside

  • Compatible gel tanks
  • Choose the appropriate gel
  • Sample preparation
  • Prepare the gel tank
  • Sample loading
  • Running conditions
  • Open the cassette and stain the gel

Specifications

Available gel percentages8%, 10%, 12%, 4-12%, 8-16%, 4-20% (MOPS and MES)
Max. volume per well80 μl (10-well), 60 μl (12-well), 40 μl (15-well)
Running voltage200 V
MOPS running conditionsStarted current 95-120 mA, finished current 35-55 mA, 30+ minutes per gel
MES running conditionsStarted current 110-130 mA, finished current 50-70 mA, 20+ minutes per gel

Frequently asked questions

Which gel tanks are compatible with SurePage Gels 10×8?

The guide lists Bio-Rad Mini-Protean (II/3/Tetra System), Hoefer Mighty Small (SE 250/SE 260), and Invitrogen Novex XCell I, II & SureLock tanks; see the manual or website for the complete list.

How should samples be prepared before loading?

Heat the sample at 70°C for 10 min and spin it down before loading; 4×LDS sample buffer (Cat. No. M00676) is recommended.

Which running buffer should be used?

Use MOPS (Cat. No. M00138) or MES buffer (Cat. No. M00677) as the running buffer.

How do I remove the gel from the cassette?

Open the cassette with GenScript's opener or a screw driver, then stain the gel, for example with the eStain L1 automatic Protein Staining System.