GenScript guide for using SurePage Gels 10×8 in protein electrophoresis, covering compatible gel tanks, sample prep, loading, and running conditions.
This guide explains how to use SurePage™ Gels, 10×8, precast gels for protein electrophoresis from GenScript. It lists compatible gel tanks, including the Bio-Rad Mini-Protean (II/3/Tetra System), Hoefer Mighty Small (SE 250/SE 260), and Invitrogen Novex XCell I, II & SureLock, and covers both MOPS and MES buffer systems.
The numbered steps cover choosing a gel percentage, preparing samples with 4×LDS sample buffer, preparing the gel tank, peeling the tape and removing the comb, loading samples, running conditions at 200 V, and opening the cassette to remove and stain the gel. GenScript support contact details are included at the end.
| Available gel percentages | 8%, 10%, 12%, 4-12%, 8-16%, 4-20% (MOPS and MES) |
|---|---|
| Max. volume per well | 80 μl (10-well), 60 μl (12-well), 40 μl (15-well) |
| Running voltage | 200 V |
| MOPS running conditions | Started current 95-120 mA, finished current 35-55 mA, 30+ minutes per gel |
| MES running conditions | Started current 110-130 mA, finished current 50-70 mA, 20+ minutes per gel |
The guide lists Bio-Rad Mini-Protean (II/3/Tetra System), Hoefer Mighty Small (SE 250/SE 260), and Invitrogen Novex XCell I, II & SureLock tanks; see the manual or website for the complete list.
Heat the sample at 70°C for 10 min and spin it down before loading; 4×LDS sample buffer (Cat. No. M00676) is recommended.
Use MOPS (Cat. No. M00138) or MES buffer (Cat. No. M00677) as the running buffer.
Open the cassette with GenScript's opener or a screw driver, then stain the gel, for example with the eStain L1 automatic Protein Staining System.