PerkinElmer SPB Graphite Digestion Instrument Operating Procedure

This manual describes the step‑by‑step procedure for soil four‑acid digestion using the PerkinElmer SPB graphite digestion instrument.


PerkinElmer SPB Graphite Digestion Instrument Operating Procedure - cover page
Brand
PerkinElmer
Category
Other
Document type
User Manual
Language
Chinese
Pages
1
File format
PDF
File size
1.2 MB
Published
28 September, 2026
Updated
28 September, 2026
MD5 checksum
1B45316402565408FC99550341705039

About this manual

The document provides a complete operating procedure for the SPB graphite digestion instrument, covering sample preparation, reagent addition, heating stages, acid evaporation, and final dilution. It is intended for laboratory personnel performing soil digestion prior to analysis by AAS, ICP‑OES, or ICP‑MS.

Key steps include weighing the soil sample, adding appropriate volumes of aqua regia, hydrofluoric acid and perchloric acid, controlling temperature ramps from 110 °C up to 175 °C, and ensuring consistent acid volumes across samples and blanks. Safety notes emphasize working in a fume hood and avoiding high‑temperature acid additions that could cause explosions. This manual is written in Chinese.

What's inside

  • Reagents and Consumables
  • Equipment
  • Sample Preparation
  • Acid Addition
  • Heating and Temperature Control
  • Acid Evaporation (Rushing)
  • Final Dilution
  • Safety Precautions

Specifications

Cover material and temperature limitPP, up to 130 °C; PTFE optional for higher temperature
Sample weight for AAS/ICP‑OES0.2500–0.5000 g
Sample weight for ICP‑MS0.1000 g
Aqua regia volume for 0.5 g sample6 mL
Hydrofluoric acid volume4 mL
Perchloric acid volume2 mL

Frequently asked questions

What sample weight should be used for AAS or ICP‑OES analysis?

Weigh 0.2500 to 0.5000 g of soil sample for AAS or ICP‑OES.

How much aqua regia is required for a 0.5 g soil sample?

Add 6 mL of aqua regia to a 0.5 g sample.

What temperature should be set after adding hydrofluoric acid?

Heat the tube at 120 °C for 4 hours after adding HF.

What material is the digestion tube cover made of and what is its temperature limit?

The cover is made of PP and can withstand up to 130 °C; PTFE can be used for higher temperatures.