The SITRANS LVL 200S, LVL 200E from Siemens is a SIL 2 certified transistor (NPN/PNP). The product is suitable for safety-related applications such as level measurement in the process industry.
This safety manual applies to Siemens SITRANS LVL 200S and LVL 200E vibration level sensors with SIL qualification and a transistor (NPN/PNP) electronic module. It covers the safety function, safe state, operating prerequisites, and safety-related technical parameters for use in safety-related systems according to IEC 61508.
It also describes commissioning adjustments, internal diagnostics and service behavior, periodic proof tests, and annexes including a test protocol, definitions, and SIL conformity. The sensors are intended for level detection in liquids in low demand or high demand mode, up to SIL2 in single-channel and up to SIL3 in multichannel architectures. This manual is written in Spanish.
| Electronic module | Transistor (NPN/PNP) |
|---|---|
| Safety Integrity Level (single-channel) | SIL2 |
| Safety Integrity Level (multichannel) | SIL3 |
| Instrument type | Type A |
| Operating mode | Low demand mode, High demand mode |
| MTBF | 3.61 x 10^6 h (412 years) |
| PFDAVG (T1 = 1 year) | 0.025 x 10^-2 |
| PFH | 0.030 x 10^-6 1/h |
| MTTFd | 3805 years |
| Performance Level | 3.00 x 10^-8 1/h (corresponds to e) |
| Safe state output | Transistors non-conductive (no current) |
| Ambient temperature correction factor | 1.3 above 50 C; 2.5 above 60 C |
| Mean time to repair | 24 h |
Up to SIL2 in a single-channel architecture and up to SIL3 in a multichannel architecture.
The transistors are non-conductive (no current).
Take the entire measuring system out of service and keep the process safe by other measures; report undetected dangerous faults to the manufacturer.
One test without filling/emptying or removing the sensor and one with filling/emptying or removing the sensor.