The MP 120 is a pressure meter that has the following features: easy to use, adjustable backlight, hold-min-max function. It can measure pressure, air velocity, coefficient of the adjustable differential pressure element, temperature and atmospheric pressure. It can select units, manual autozero, hold function, display minimum and maximum values, configurable auto shut-off and backlight.
The MP 120 is a pressure measuring instrument built around a piezoresistive sensor. It measures pressure and air velocity and offers a hold-min-max function, manual autozero, unit selection, an adjustable backlight and a configurable auto shut-off. Pressure is displayed on a 4-line LCD with 5-digit value lines and 5-digit unit lines.
This leaflet lists the instrument's functions, technical features and limits: an allowed overpressure of 250 mbar, a pressure range of -1000 to +1000 Pa, a velocity range of 0 to 40 m/s, IP54 housing protection and 180 hours of battery life on four AAA cells. It also explains the piezoresistive operating principle, the Pitot tube method used for air velocity, the supplied accessories and the yearly checking recommendation.
| Measuring element | Piezoresistive sensor |
|---|---|
| Overpressure allowed | 250 mbar |
| Pressure measuring range | -1000 to +1000 Pa |
| Pressure accuracy | ±0.5% of reading ±2 Pa, resolution 1 Pa |
| Velocity measuring range | 0 to 40 m/s |
| Display | 4 lines, LCD, 50 x 36 mm |
| Housing | ABS, protection IP54 |
| Connector | Ø 6.2 mm threaded connectors, nickelled brass |
| Power supply | 4 batteries AAA LR03 1.5 V |
| Battery life | 180 hours |
| Auto shut-off | Adjustable from 0 to 120 min |
| Storage temperature | -20 to +80 °C |
| Weight | 220 g |
It measures from -1000 to +1000 Pa, with an accuracy of ±0.5% of reading ±2 Pa and a resolution of 1 Pa.
The instrument comes with a calibration certificate, two 1 m silicone tubes Ø 4 x 7 mm, a stainless steel tip Ø 6 x 100 mm and a transport case (ref. ST 110).
Dynamic pressure is measured by a Pitot tube and air velocity is calculated with the simplified Bernoulli formula including temperature correction.
As part of Quality Assurance Standards, a yearly checking is recommended; calibration, adjustment and maintenance are carried out by the manufacturer's service.