Bacharach 0017-7000 Series Manometer Gas Pressure Gauge Operation and Maintenance Manual

This manual provides operation and maintenance instructions for Bacharach manometer gas pressure gauges (models 0017-7000, 0017-7002, 0017-7008, 0017-7013…


Bacharach 0017-7000 Series Manometer Gas Pressure Gauge Operation and Maintenance Manual - cover page
Brand
Bacharach
Category
pressure transmitters
Document type
User Manual
Language
English
Pages
7
File format
PDF
File size
85 KB
Published
07 November, 2026
Updated
07 November, 2026
MD5 checksum
17856CA15C5B9B5C2EA374ACE2183B36

About this manual

The document covers Bacharach manometer gas pressure gauges with reversible scales for measuring inches of water, centimeters of water, and ounces per square inch. It includes warranty information, operating instructions for connecting to a gas line and reading manifold pressure, application guidance for determining firing rates, and maintenance procedures such as reversing the scale, adding indicating fluid, and replacing parts.

A detailed parts list with part numbers and descriptions is provided for all models, along with an exploded view diagram. The manual specifies using only Bacharach Fluid 0017-0025 and warns that water will not operate in certain models due to blow-over seals.

What's inside

  • Warranty
  • Operating Instructions
  • Connecting to Gas Line
  • Reading-Adjusting Manifold Pressure
  • Application Instructions
  • Maintenance Instructions
  • Reversing Scale
  • Adding Indicating Fluid
  • Parts Replacement
  • Parts List

Specifications

Range (0017-7000/7008)0-15 inches of water and 38 cm water
Range (0017-7002)0-15 in. of water and 8.5 oz./sq. in. water
Range (0017-7013/7016)0-7 inches of water and 18 cm water
Scale typeReversible
Indicating fluidBacharach Fluid 0017-0025
Warranty period1 year from delivery

Frequently asked questions

How do I read the manifold pressure on the gauge?

Slide the scale up until the zero scale arrow is opposite the top of the low (right-hand) liquid column, then read the gas manifold pressure on the scale opposite the top of the higher (left-hand) liquid column. This is a direct reading; do not multiply by 2.

How do I reverse the scale on model 0017-7002 to read ounces per square inch?

Close both valves, remove the screw, press the retaining ring out of the grooves, lift off the front cover, remove the scale knob, scale clamp cover, and scale clamp screw, reverse the scale, and reassemble.